| Literature DB >> 18094804 |
Kilza de Arruda Lyra e Silva1, Beatriz de Albuquerque C C Novaes, Dóris Ruthi Lewis, Renata Mota Mamede Carvallo.
Abstract
UNLABELLED: Tympanometry is used in evaluating middle ear functional conditions. Before six months of age its results may be misleading. High frequency studies aim to provide more valid procedures. AIM: To describe and discuss tympanometric measurements and the interpretation in normal hearing neonates at 226, 678 and 1000Hz.Entities:
Mesh:
Year: 2007 PMID: 18094804 PMCID: PMC9445730 DOI: 10.1016/s1808-8694(15)30123-3
Source DB: PubMed Journal: Braz J Otorhinolaryngol ISSN: 1808-8686
Tympanometric curves related to probe frequency in 110 neonates (58 male and 52 female) in both ears.
| Type | 226 Hz | 678 Hz | 1000 Hz | |||
|---|---|---|---|---|---|---|
| N | % | N | % | N | % | |
| PU | 105 | 47,7 | 56 | 25,4 | 156 | 70,9 |
| PD | 115 | 52,3 | ||||
| AS | 148 | 67,3 | 62 | 28,2 | ||
| I | 16 | 7,3 | 2 | 0,9 | ||
| Total | 220 | 100,0 | 220 | 100,0 | 220 | 100,0 |
Total equal and different curve types in both ears in 110 male and female neonates.
| Type | 226 Hz | 678 Hz | 1000 Hz | |||
|---|---|---|---|---|---|---|
| N | % | N | % | N | % | |
| Equal | ||||||
| PU | 36 | 32,7 | 16 | 14,5 | 68 | 61,8 |
| PD | 41 | 37,3 | ||||
| AS | 60 | 54,5 | 21 | 19,1 | ||
| I | 2 | 1,8 | 1 | 0,9 | ||
| Different | 33 | 30,0 | 32 | 29,1 | 20 | 18,2 |
| Total | 110 | 100,0 | 110 | 100,0 | 110 | 100,0 |
Joint comparison of curve types for 226 and 1000 Hz probes in both ears in 110 male and female neonates.
| 226 Hz / 1000 Hz | N | % |
|---|---|---|
| PU / PU | 23 | 20,9 |
| PD / PU | 26 | 23,6 |
| Other | 61 | 55,5 |
| Total | 110 | 100,0 |
Results of the statistical analysis - p-values obtained by ANOVA.
| Probe | Variable | Equality of means test in both ears | Equality of means test in both sexes |
|---|---|---|---|
| p-value | p-value | ||
| 226 Hz | Vea | 0,100 | 0,026 |
| TPP | 0,538 | 0,716 | |
| Ymt | 0,080 | 0,144 | |
| TW | 0,058 | 0,989 | |
| 1000 Hz | TPP | 0,392 | 0,690 |
| Ymt | 0,924 | 0,000 | |
| TW | 0,817 | 0,958 |
Quantitative measurements of 220 tympanograms with the 226 Hz probe.
| Measurement | Sex | N | Mean | Standard deviation | 5 | Percentiles | 95 |
|---|---|---|---|---|---|---|---|
| Vea | Female | 104 | 0,64 | 0,01 | 0,50 | 0,60 | 0,80 |
| Male | 116 | 0,68 | 0,01 | 0,50 | 0,60 | 0,82 | |
| TPP | 220 | -8,93 | 31,40 | -60,00 | -5,00 | 39,75 | |
| Ymt | 220 | 0,75 | 0,22 | 0,40 | 0,80 | 1,10 | |
| TW | 220 | 115,05 | 40,60 | 45,25 | 115,00 | 184,75 | |
Quantitative measurements of 220 tympanograms with the 1000 Hz probe.
| Measurement | Sex | N | Mean | Standard deviation | 5 | Percentiles | 95 |
|---|---|---|---|---|---|---|---|
| TPP | 156 | -3,46 | 44,21 | -90,00 | 0,00 | 61,50 | |
| Ymt | Female | 72 | 0,90 | 0,47 | 0,30 | 0,75 | 1,94 |
| Male | 84 | 1,36 | 0,75 | 0,50 | 1,20 | 3,15 | |
| TW | 156 | 99,49 | 24,73 | 60,00 | 100,00 | 145,00 |
Result of the tympanometric analysis based on Sutton et al.’s14 protocol in 220 ears using test probe tones at 678 and 1000Hz.
| 678 Hz | 1000 Hz | |||
|---|---|---|---|---|
| N | % | N | % | |
| Normal | 208 | 94,5 | 217 | 98,6 |
| Abnormal | 12 | 4,5 | 3 | 1,4 |
| Total | 220 | 100,0 | 220 | 100,0 |