Literature DB >> 18083629

An investigation of the validity of the Lifeshirt in comparison to standard polysomnography in the detection of obstructive sleep apnea.

Suanne Goodrich1, William C Orr.   

Abstract

BACKGROUND: Due to the cost and waiting times for traditional sleep studies, there is great interest in finding alternatives for the diagnosis of sleep apnea. Several alternatives exist, including ambulatory devices. Our laboratory utilized one such device, a form-fitting vest called the Lifeshirt, and compared its accuracy to that of polysomnography (PSG).
METHODS: PSG was performed with simultaneous utilization of the Lifeshirt on 50 individuals who met screening criteria for obstructive sleep apnea. Participants came to the sleep laboratory approximately 2 h before their normal bedtime. A sleep technician prepared each participant for PSG and Lifeshirt monitoring. "Lights Out" occurred when the participant was ready for bed, and time in bed was standardized to approximately 7 h. PSGs were scored by experienced personnel in our laboratory, while Lifeshirt data were sent electronically to the Lifeshirt manufacturer for analysis. The major variable that was compared between PSG and Lifeshirt was the apnea hypopnea index (AHI), or the number of apneas and hypopneas per hour. Due to incomplete data on two participants, analyses were completed on 48 individuals.
RESULTS: Sensitivity of the Lifeshirt ranged from .85 (AHI of > or =5) to 1.00 (AHI of > or =25). Specificity ranged from .67 to 1.00. Using the Bland-Altman technique of determining agreement, the mean difference between the AHI of the Lifeshirt and PSG was 1.02 (+/-16.36). When these values are plotted, every case falls within the limits of agreement, with one exception.
CONCLUSIONS: The Lifeshirt compared favorably with PSG and could be used with considerable confidence for the screening of patients with suspected obstructive sleep apnea.

Entities:  

Mesh:

Year:  2008        PMID: 18083629     DOI: 10.1016/j.sleep.2007.11.003

Source DB:  PubMed          Journal:  Sleep Med        ISSN: 1389-9457            Impact factor:   3.492


  9 in total

1.  Auto-titrating continuous positive airway pressure for patients with acute transient ischemic attack: a randomized feasibility trial.

Authors:  Dawn M Bravata; John Concato; Terri Fried; Noshene Ranjbar; Tanesh Sadarangani; Vincent McClain; Frederick Struve; Lawrence Zygmunt; Herbert J Knight; Albert Lo; George B Richerson; Mark Gorman; Linda S Williams; Lawrence M Brass; Joseph Agostini; Vahid Mohsenin; Francoise Roux; H Klar Yaggi
Journal:  Stroke       Date:  2010-05-27       Impact factor: 7.914

2.  Continuous positive airway pressure: evaluation of a novel therapy for patients with acute ischemic stroke.

Authors:  Dawn M Bravata; John Concato; Terri Fried; Noshene Ranjbar; Tanesh Sadarangani; Vincent McClain; Frederick Struve; Lawrence Zygmunt; Herbert J Knight; Albert Lo; George B Richerson; Mark Gorman; Linda S Williams; Lawrence M Brass; Joseph Agostini; Vahid Mohsenin; Francoise Roux; H Klar Yaggi
Journal:  Sleep       Date:  2011-09-01       Impact factor: 5.849

3.  Laboratory Validation of Hexoskin Biometric Shirt at Rest, Submaximal Exercise, and Maximal Exercise While Riding a Stationary Bicycle.

Authors:  Cara M Smith; Steven N Chillrud; Darby W Jack; Patrick Kinney; Qiang Yang; Aimee M Layton
Journal:  J Occup Environ Med       Date:  2019-04       Impact factor: 2.162

4.  Sleep apnea monitoring and diagnosis based on pulse oximetry and tracheal sound signals.

Authors:  Azadeh Yadollahi; Eleni Giannouli; Zahra Moussavi
Journal:  Med Biol Eng Comput       Date:  2010-08-24       Impact factor: 2.602

5.  Design and evaluation of a ubiquitous chest-worn cardiopulmonary monitoring system for healthcare application: a pilot study.

Authors:  Jiewen Zheng; Congying Ha; Zhengbo Zhang
Journal:  Med Biol Eng Comput       Date:  2016-05-13       Impact factor: 2.602

Review 6.  Diagnostic accuracy of level 3 portable sleep tests versus level 1 polysomnography for sleep-disordered breathing: a systematic review and meta-analysis.

Authors:  Mohamed El Shayeb; Leigh-Ann Topfer; Tania Stafinski; Lawrence Pawluk; Devidas Menon
Journal:  CMAJ       Date:  2013-11-11       Impact factor: 8.262

Review 7.  Computer-Assisted Diagnosis of the Sleep Apnea-Hypopnea Syndrome: A Review.

Authors:  Diego Alvarez-Estevez; Vicente Moret-Bonillo
Journal:  Sleep Disord       Date:  2015-07-21

8.  Theory and developments in an unobtrusive cardiovascular system representation: ballistocardiography.

Authors:  Eduardo Pinheiro; Octavian Postolache; Pedro Girão
Journal:  Open Biomed Eng J       Date:  2010-10-10

9.  Utility of a smartphone based system (cvrphone) to accurately determine apneic events from electrocardiographic signals.

Authors:  Kwanghyun Sohn; Faisal M Merchant; Shady Abohashem; Kanchan Kulkarni; Jagmeet P Singh; E Kevin Heist; Chris Owen; Jesse D Roberts; Eric M Isselbacher; Furrukh Sana; Antonis A Armoundas
Journal:  PLoS One       Date:  2019-06-17       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.