Literature DB >> 18082631

Undetectable phenytoin serum levels by an automated particle-enhanced turbidimetric inhibition immunoassay in a patient with monoclonal IgM lambda.

Yolanda B Brauchli1, André Scholer, Martin Schwietert, Stephan Krähenbühl.   

Abstract

INTRODUCTION: Phenytoin is a drug used for the treatment of different types of seizures. Its variable pharmacokinetics, mainly a consequence of variable bioavailability, saturable protein binding and saturable hepatic metabolism, predisposes the drug to therapeutic drug monitoring. Several methods to analyze the drug in serum exist with immunoassays being the method of choice for routine measurements. Immunoassays are specific and sensitive, but cross-reactivity, possibly leading to erroneous serum levels, is a concern. We report a patient with falsely undetectable phenytoin serum levels. CASE REPORT: This 73-year old woman was treated with intravenous phenytoin due to epilepsia partialis continua in the context of a bilateral cerebrovascular insult. Anamnestically, a chronic lymphatic leukemia was known. In this patient, serum phenytoin levels became only detectable by the particle enhanced turbidimetric inhibition immunoassay used after precipitation of serum proteins. Protein electrophoresis revealed a monoclonal immunoglobulin, identified as IgMlambda. With other methods such as HPLC and fluorescence depolarization immunoassay, phenytoin was detectable.
CONCLUSION: We propose interference between the monoclonal IgMlambda and/or other serum proteins and the particle-enhanced turbidimetric inhibition immunoassay, rendering phenytoin falsely undetectable in samples of this patient. In such patients, alternative methods such as HPLC should be used to prevent dosage errors.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18082631     DOI: 10.1016/j.cca.2007.11.014

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  2 in total

1.  Paraprotein interference with turbidimetric gentamicin assay.

Authors:  Goce Dimeski; Kendra Bassett; Nigel Brown
Journal:  Biochem Med (Zagreb)       Date:  2015       Impact factor: 2.313

2.  Comparison of performance characteristics between high-performance liquid chromatography and latex agglutination turbidimetric immunoassay for therapeutic drug monitoring of zonisamide.

Authors:  Daiki Eto; Ryota Tanaka; Yosuke Suzuki; Yuhki Sato; Hiroki Itoh
Journal:  J Clin Lab Anal       Date:  2019-06-20       Impact factor: 2.352