Christopher L Brace1, Daniel W van der Weide, Fred T Lee, Paul F Laeseke, Lisa Sampson.
Abstract
We apply a new triaxial antenna for microwave ablation procedures. The antenna consists of a coaxial monopole inserted through an 18-gauge biopsy needle positioned one quarter-wavelength from the antenna base. The biopsy needle creates a triaxial structure, which enhances return loss by more than 10 dB, thus limiting return currents along the feed line. Numerical simulations are used to optimize the antenna design. Numerical and ex-vivo experimental results are presented to quantify the field distribution, heating pattern and return loss of the antenna.Entities:
Year: 2004 PMID: 18079982 PMCID: PMC2134895 DOI: 10.1109/MWSYM.2004.1338842