| Literature DB >> 18052498 |
Hakki Yegingil1, Wan Y Shih, Wei-Heng Shih.
Abstract
We have experimentally investigated the depth sensitivity limit of a piezoelectric cantilever tissue elastic modulus sensor and simultaneously determined the elastic modulus and the depth of a tumor directly. Using model tissues consisting of bottom-supported modeling clay inclusions of various depths in a gelatin matrix, we empirically determined that the depth sensitivity limit of a piezoelectric cantilever sensor was twice the linear dimension of the indentation area (or the cantilever width). Knowing the depth sensitivity limit of the individual cantilever sensor as input and treating a model tissue that has the gelatin matrix on top and the modeling clay inclusion at the bottom as two springs in series, we showed that the elastic moduli and depths of the hard inclusions could be simultaneously determined with the elastic modulus profiles measured by two cantilevers with different widths as input.Entities:
Mesh:
Year: 2007 PMID: 18052498 DOI: 10.1063/1.2793502
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523