Literature DB >> 18042856

An experimental and clinical evaluation of a novel central venous catheter with integrated oximetry for pediatric patients undergoing cardiac surgery.

Oliver J Liakopoulos1, Jonathan K Ho, Aaron Yezbick, Elizabeth Sanchez, Clayton Naddell, Gerald D Buckberg, Ryan Crowley, Aman Mahajan.   

Abstract

BACKGROUND: Central venous oxygen saturation (ScvO2) accurately reflects cardiocirculatory function, but is not always feasible in pediatric patients. Using an experimental and clinical approach, we determined the accuracy of a novel pediatric central venous catheter with integrated fiberoptic oximetry, correlated ScvO2 to periprocedural vital variables, and tested its feasibility in pediatric cardiac surgery patients.
METHODS: In five anesthetized pigs, hemodynamics (cardiac index [CI], heart rate; mean arterial blood [MAP]; mean pulmonary artery [MPAP], central venous pressure [CVP]), fiberoptic ScvO2 (ScvO2-cath), and blood gas oximetry (ScvO2-blood) were measured during stable baseline conditions, preload reduction (caval occlusion), and dopamine infusion (5 mcg x kg(-1) x min(-1)). In 16 pediatric patients undergoing cardiac surgery (median age 8.4 mo; weight 8.0 kg), central venous oximetry catheters were placed percutaneously, and ScvO2-cath and hemodynamics recorded at several time-points during and until 24 h after surgery. Oximetry and hemodynamic data were compared by correlation (Pr) and the Bland-Altman analysis.
RESULTS: There were no catheter-related complications. ScvO2-cath and ScvO2-blood measurements correlated significantly (P < 0.001) in both the experimental (Pr = 0.96) and clinical protocol (Pr = 0.94). A similar bias and precision over all time-points was detected in both protocols (Exp-bias: +0.03% +/- 4.11%; Clinical-bias: -0.03% +/- 4.41%). ScvO2-cath correlated (P < 0.001) with CI (Pr = 0.87), MAP (Pr = 0.59), MPAP (Pr = 0.44), and CVP (Pr = 0.38) and estimated CI better than MAP (Pr = 0.61), MPAP (Pr = 0.38), CVP (Pr = 0.35), or heart rate (Pr = 0.25).
CONCLUSION: Integrated central venous oximetry catheters provide accurate continuous ScvO2 monitoring in pediatric patients undergoing cardiac surgery. ScvO2 fiberoptic oximetry correlates better with changes in CI as compared to routine hemodynamic variables.

Entities:  

Mesh:

Year:  2007        PMID: 18042856     DOI: 10.1213/01.ane.0000287657.08434.dc

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  8 in total

1.  Inferior Vena Cava Oxygen Saturation during the First Three Postnatal Days in Preterm Newborns with and without Patent Ductus Arteriosus.

Authors:  Ece Yapakçı; Ayşe Ecevit; Deniz Anuk Ince; Mahmut Gökdemir; M Agah Tekindal; Hande Gülcan; Aylin Tarcan
Journal:  Balkan Med J       Date:  2014-09-01       Impact factor: 2.021

2.  Wireless implantable optical probe for continuous monitoring of oxygen saturation in flaps and organ grafts.

Authors:  Hexia Guo; Wubin Bai; Wei Ouyang; Yihan Liu; Changsheng Wu; Yameng Xu; Yang Weng; Hao Zang; Yiming Liu; Lauren Jacobson; Ziying Hu; Yihang Wang; Hany M Arafa; Quansan Yang; Di Lu; Shuo Li; Lin Zhang; Xun Xiao; Abraham Vázquez-Guardado; Joanna Ciatti; Elizabeth Dempsey; Nayereh Ghoreishi-Haack; Emily A Waters; Chad R Haney; Amanda M Westman; Matthew R MacEwan; Mitchell A Pet; John A Rogers
Journal:  Nat Commun       Date:  2022-05-30       Impact factor: 17.694

3.  Use of central venous saturation monitoring in a patient with pediatric cardiac beriberi.

Authors:  Nozomi Majima; Osamu Umegaki; Masako Soen
Journal:  World J Clin Cases       Date:  2013-09-16       Impact factor: 1.337

4.  Computer-controlled closed-loop drug infusion system for automated hemodynamic resuscitation in endotoxin-induced shock.

Authors:  Kazunori Uemura; Toru Kawada; Can Zheng; Meihua Li; Masaru Sugimachi
Journal:  BMC Anesthesiol       Date:  2017-10-23       Impact factor: 2.217

5.  A validation method for near-infrared spectroscopy based tissue oximeters for cerebral and somatic tissue oxygen saturation measurements.

Authors:  Paul B Benni; David MacLeod; Keita Ikeda; Hung-Mo Lin
Journal:  J Clin Monit Comput       Date:  2017-04-03       Impact factor: 2.502

6.  Wireless, implantable catheter-type oximeter designed for cardiac oxygen saturation.

Authors:  Wei Lu; Wubin Bai; Hao Zhang; Chenkai Xu; Antonio M Chiarelli; Abraham Vázquez-Guardado; Zhaoqian Xie; Haixu Shen; Khizar Nandoliya; Hangbo Zhao; KunHyuck Lee; Yixin Wu; Daniel Franklin; Raudel Avila; Shuai Xu; Alina Rwei; Mengdi Han; Kyeongha Kwon; Yujun Deng; Xinge Yu; Edward B Thorp; Xue Feng; Yonggang Huang; Joseph Forbess; Zhi-Dong Ge; John A Rogers
Journal:  Sci Adv       Date:  2021-02-10       Impact factor: 14.136

7.  Central Venous-to-Arterial CO2 Gap Is a Useful Parameter in Monitoring Hypovolemia-Caused Altered Oxygen Balance: Animal Study.

Authors:  Szilvia Kocsi; Gabor Demeter; Daniel Erces; Eniko Nagy; Jozsef Kaszaki; Zsolt Molnar
Journal:  Crit Care Res Pract       Date:  2013-08-29

Review 8.  Biochemical markers for clinical monitoring of tissue perfusion.

Authors:  Marek Janotka; Petr Ostadal
Journal:  Mol Cell Biochem       Date:  2021-01-02       Impact factor: 3.396

  8 in total