Literature DB >> 18038874

The variability of applicator position among high dose rate intracavitary brachytherapy applications in cervical cancer patients treated with ring & tandem applicators.

Cuneyt Ebruli1, Ayşe Nur Demiral, Riza Cetingöz, Ferhat Eyiler, Münir Kinay.   

Abstract

OBJECTIVE: To evaluate the interindividual and intraindividual applicator position variability in high dose rate ring and tandem intracavitary brachytherapy applications in locally advanced cervical cancer. PATIENTS AND METHODS: Eight patients with locally advanced cervical cancer formed the study population. They had been treated in Dokuz Eylül University Department of Radiation Oncology between the years 2000 and 2005 with high dose rate intracavitary brachytherapy using ring and tandem applicators. The 3-dimensional geometric variation of the applicator center in craniocaudal, mediolateral and anteroposterior directions was determined on the basis of bony reference points in 24 pairs of orthogonal films obtained in the conventional simulator. Then the following evaluations were performed: 1) the applicator position variability in all applications (interindividual variability), 2) the intraindividual applicator position variability relative to the first application, 3) the intraindividual applicator position variability relative to the average of three applications. Among the potential factors that might influence the reproducibility of ring and tandem applications, age, stage, the period between external radiotherapy and brachytherapy were evaluated by univariate analysis.
RESULTS: Standard deviation of interindividual applicator variability was 3.83 mm in craniocaudal, 0.39 mm in mediolateral and 2.86 mm in anteroposterior directions. The standard deviation of intraindividual variability relative to the first application was 1.91 mm in craniocaudal, 0.4 mm in mediolateral, and 4.26 mm in anteroposterior directions. The standard deviation of intraindividual variability relative to the average of three applications was 0.95 mm in craniocaudal, 1.86 mm in mediolateral, and 1.24 mm in anteroposterior directions. According to univariate analysis, no factor influenced applicator position variability.
CONCLUSIONS: In order to extract definitive conclusions about factors that affect positional reproducibility of ring and tandem applicators, studies are needed that include larger numbers of patients.

Entities:  

Mesh:

Year:  2007        PMID: 18038874     DOI: 10.1177/030089160709300505

Source DB:  PubMed          Journal:  Tumori        ISSN: 0300-8916


  6 in total

1.  The impact of maximum rectal distention and tandem angle on rectal dose delivered in 3D planned gynecologic high dose-rate brachytherapy.

Authors:  Jihoon Lim; Blythe Durbin-Johnson; Richard Valicenti; Matthew Mathai; Robin L Stern; Jyoti Mayadev
Journal:  Int J Gynecol Cancer       Date:  2013-07       Impact factor: 3.437

2.  Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets.

Authors:  Ryan McMahon; Tingliang Zhuang; Beverly A Steffey; Haijun Song; Oana I Craciunescu
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

3.  Dosimetric evaluation of image based brachytherapy using tandem ovoid and tandem ring applicators.

Authors:  Ramya Rangarajan
Journal:  Rep Pract Oncol Radiother       Date:  2018-01-12

4.  A comparison of dose distribution from Manchester-style and Fletcher-style intracavitary brachytherapy applicator systems in cervical cancer.

Authors:  Bishan Basu; Swapnendu Basu; Bikramjit Chakraborti; Suman Ghorai; Phalguni Gupta; Sajal Ghosh; Koushik Ghosh; J Jayanti
Journal:  J Contemp Brachytherapy       Date:  2012-12-28

5.  Magnitude and Implications of Interfraction Variations in Organ Doses during High Dose Rate Brachytherapy of Cervix Cancer: A CT Based Planning Study.

Authors:  Santam Chakraborty; Firuza D Patel; Vijay M Patil; Arun S Oinam; Suresh C Sharma
Journal:  ISRN Oncol       Date:  2014-02-03

6.  Comparison of the Dosimetric Influence of Applicator Displacement on 2D and 3D Brachytherapy for Cervical Cancer Treatment.

Authors:  Ailin Wu; Du Tang; Aidong Wu; Yunqin Liu; Liting Qian; Lei Zhu
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.