Literature DB >> 18021823

In vitro model to evaluate reliability and accuracy of a dental shade-matching instrument.

Seungyee Kim-Pusateri1, Jane D Brewer, Robert G Dunford, Alvin G Wee.   

Abstract

STATEMENT OF PROBLEM: There are several electronic shade-matching instruments available for clinical use; unfortunately, there are limited acceptable in vitro models to evaluate their reliability and accuracy.
PURPOSE: The purpose of this in vitro study was to evaluate the reliability and accuracy of a dental clinical shade-matching instrument.
MATERIAL AND METHODS: Using the shade-matching instrument (ShadeScan), color measurements were made of 3 commercial shade guides (VITA Classical, VITA 3D-Master, and Chromascop). Shade tabs were selected and placed in the middle of a gingival matrix (Shofu Gummy), with tabs of the same nominal shade from additional shade guides placed on both sides. Measurements were made of the central region of the shade tab inside a black box. For the reliability assessment, each shade tab from each of the 3 shade guide types was measured 10 times. For the accuracy assessment, each shade tab from 10 guides of each of the 3 types evaluated was measured once. Reliability, accuracy, and 95% confidence intervals were calculated for each shade tab. Differences were determined by 1-way ANOVA followed by the Bonferroni multiple comparison procedure.
RESULTS: Reliability of ShadeScan was as follows: VITA Classical = 95.0%, VITA 3D-Master = 91.2%, and Chromascop = 76.5%. Accuracy of ShadeScan was as follows: VITA Classical = 65.0%, VITA 3D-Master = 54.2%, Chromascop = 84.5%.
CONCLUSIONS: This in vitro study showed a varying degree of reliability and accuracy for ShadeScan, depending on the type of shade guide system used.

Mesh:

Year:  2007        PMID: 18021823     DOI: 10.1016/S0022-3913(07)60119-X

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  6 in total

1.  In Vivo Evaluations of Inter-Observer Reliability Using VITA Easyshade® Advance 4.0 Dental Shade-Matching Device.

Authors:  Dubravka Knezović; Davor Zlatarić; Iva Ž Illeš; Maja Alajbeg
Journal:  Acta Stomatol Croat       Date:  2016-03

2.  In Vivo and in Vitro Evaluations of Repeatability and Accuracy of VITA Easyshade® Advance 4.0 Dental Shade-Matching Device.

Authors:  Dubravka Knezović; Davor Zlatarić; Iva Ž Illeš; Maja Alajbeg
Journal:  Acta Stomatol Croat       Date:  2015-06

3.  Color stability of hybrid ceramics exposed to beverages in different combinations.

Authors:  Kaan Yerliyurt; Işıl Sarıkaya
Journal:  BMC Oral Health       Date:  2022-05-14       Impact factor: 3.747

4.  Evaluation of bleaching efficacy of 37.5% hydrogen peroxide on human teeth using different modes of activations: An in vitro study.

Authors:  Neha Bhutani; Bhuvan Shome Venigalla; Jaya Prakash Patil; Thakur Veerandar Singh; Sistla Venkata Jyotsna; Abhilasha Jain
Journal:  J Conserv Dent       Date:  2016 May-Jun

5.  Assessing the effect of multibracket appliance treatment on tooth color by using electronic measurement.

Authors:  Anja Ratzmann; Christan Schwahn; Anja Treichel; Andreas Faltermeier; Alexander Welk
Journal:  Head Face Med       Date:  2018-10-22       Impact factor: 2.151

6.  Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: an in vivo and in vitro study.

Authors:  Hee-Kyung Kim
Journal:  J Adv Prosthodont       Date:  2018-06-12       Impact factor: 1.904

  6 in total