Reliability assessment of glucose measurement by HemoCue analyser in a neonatal intensive care unit.

Carlo Bellini1, Giovanni Serra, Domenico Risso, Massimo Mazzella, Eugenio Bonioli.   

Abstract

BACKGROUND: Rapid and reliable bed-side determination of blood glucose concentration is very important in the management of acutely ill infants and especially in premature newborns. HemoCue is an easy-to-use glucose analyser. The aim of the present study was to examine the usefulness of the HemoCue glucose analyser compared to a reference plasma glucose method (SYS, BM/Hitachi 747/737) in a neonatal intensive care unit (NICU).
METHODS: Seventy-eight consecutive neonates admitted to our NICU were enrolled in the study. At the time of the study all patients were grouped according to nutritional management (parenteral or enteral nutrition), haematocrit values and birth weight. The effects of feeding management, haematocrit values, and birth weight on accuracy and precision of the device were evaluated.
RESULTS: Overall data linear regression analysis yielded an r-value of 0.905 and the Bland-Altman method demonstrated that HemoCue overestimates plasma glucose by 0.932 mmol/L. Evaluation of our data by receiver operating characteristic curve demonstrated 100% sensitivity cutoff at 4.1 mmol/L.
CONCLUSIONS: HemoCue cannot be used satisfactorily in the management of glycaemia in the NICU. In the preterm population, birth weight had a dramatic influence on HemoCue accuracy. Low haematocrit and parenteral feeding further contributed to a decrease in the accuracy of this device.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17970709     DOI: 10.1515/CCLM.2007.302

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  5 in total

Review 1.  Hypoglycemia in critically ill children.

Authors:  E Vincent S Faustino; Eliotte L Hirshberg; Clifford W Bogue
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

2.  Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting.

Authors:  Matthias Roth-Kleiner; Corinne Stadelmann Diaw; Jocelyne Urfer; Christiane Ruffieux; Dominique Werner
Journal:  Eur J Pediatr       Date:  2010-06-24       Impact factor: 3.183

Review 3.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

4.  Comparison of Point of Care (POC) Testing of Glucose by B Braun Glucometer and Hemocue Glucose 201+ Analyser Versus Centralised Testing in Neonatal Intensive Care Unit (NICU).

Authors:  Sudha Reddy V R; Sumathi M E; Beere Gowda Y C; Mohamed Suhail S
Journal:  J Clin Diagn Res       Date:  2014-07-20

5.  Cost Analysis of Cot-Side Screening Methods for Neonatal Hypoglycaemia.

Authors:  Matthew J Glasgow; Jane E Harding; Richard Edlin
Journal:  Neonatology       Date:  2018-06-12       Impact factor: 4.035