Experimental analysis of scattered photons in Tc-99m imaging with a gamma camera.

A Kojima1, M Matsumoto, M Takahashi.   

Abstract

The amount of scattered photons in a clinical imaging window of Tc-99m was experimentally measured by means of a line source with scattering materials and a gamma camera. A symmetrical photopeak energy window centered at 140 keV with a width of 20% (126-154 keV) was partitioned into several small windows. Energy spectra were analyzed to determine the scatter fraction and the attenuation coefficient for each window. Line spread functions (LSF) were also obtained to characterize the spatial scatter distribution. The results of analysis of energy spectra show that scattered photons are included over the symmetric 20% window (SW) and scatter fractions increase linearly with the increasing thickness of the scattering material in all energy windows investigated. In addition, the results for the LSF show that the scatter distribution within the SW is represented as a mono-exponential function. Experimental measurements obtained with a phantom and a gamma camera simplify accurate quantification of scattered photons. Such quantitative analysis of scattered photons is important in developing and evaluating a scatter correction technique.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1797068     DOI: 10.1007/BF03164628

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  6 in total

1.  [Estimation of scatter component in SPECT planar image using a Monte Carlo method].

Authors:  K Ogawa; Y Harata; T Ichihara; A Kubo; S Hashimoto
Journal:  Kaku Igaku       Date:  1990-05

2.  The measurement of the depth of a point source of a radioisotope from gamma ray spectra.

Authors:  L J Filipow; D J Macey; T R Munro
Journal:  Phys Med Biol       Date:  1979-03       Impact factor: 3.609

3.  Deconvolution of Compton scatter in SPECT.

Authors:  C E Floyd; R J Jaszczak; K L Greer; R E Coleman
Journal:  J Nucl Med       Date:  1985-04       Impact factor: 10.057

4.  Improved SPECT quantification using compensation for scattered photons.

Authors:  R J Jaszczak; K L Greer; C E Floyd; C C Harris; R E Coleman
Journal:  J Nucl Med       Date:  1984-08       Impact factor: 10.057

5.  Energy and spatial distribution of multiple order Compton scatter in SPECT: a Monte Carlo investigation.

Authors:  C E Floyd; R J Jaszczak; C C Harris; R E Coleman
Journal:  Phys Med Biol       Date:  1984-10       Impact factor: 3.609

6.  Subtraction of Compton-scattered photons in single-photon emission computerized tomography.

Authors:  B Axelsson; P Msaki; A Israelsson
Journal:  J Nucl Med       Date:  1984-04       Impact factor: 10.057

  6 in total
  4 in total

1.  Correction of scattered photons in Tc-99m imaging by means of a photopeak dual-energy window acquisition.

Authors:  A Kojima; A Tsuji; Y Takaki; S Tomiguchi; M Hara; M Matsumoto; M Takahashi
Journal:  Ann Nucl Med       Date:  1992-08       Impact factor: 2.668

2.  Photon energy readings in OSL dosimeter filters: an application to retrospective dose estimation for nuclear medicine workers.

Authors:  Daphnée Villoing; Cari M Kitahara; Christopher Passmore; Steven L Simon; R Craig Yoder
Journal:  J Radiol Prot       Date:  2018-06-19       Impact factor: 1.394

3.  Evaluation of quantitative accuracy among different scatter corrections for quantitative bone SPECT/CT imaging.

Authors:  Kenta Miwa; Reo Nemoto; Hirotsugu Masuko; Tensho Yamao; Rinya Kobayashi; Noriaki Miyaji; Kosuke Inoue; Hiroya Onodera
Journal:  PLoS One       Date:  2022-06-06       Impact factor: 3.752

4.  Small-angle Compton Scatter Artifact in Tc-99m-IDA Hepatobiliary Scintigraphy Resulting in the Breast Overlying the Liver in Planar Dynamic Imaging.

Authors:  Mohsen Qutbi
Journal:  Mol Imaging Radionucl Ther       Date:  2021-02-09
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.