Rapid pre-tension loss in the Ilizarov external fixator: an in vitro study.

René Aquarius1, Albert Van Kampen, Nico Verdonschot.   

Abstract

BACKGROUND AND
PURPOSE: Wire pre-tension in the Ilizarov frame is considered to be important in order to reduce movements that can impair fracture healing. Wires will eventually lose part of their pre-tension, however. In order to gain more insight into the need for wire pre-tension, we investigated: (1) the amount of pre-ten-sion loss, (2) the relationship between pre-tension loss and bolt-tightening torque, (3) the cause of a possible loss of pre-tension, and (4) the effect of pre-tension on cyclic micromotions and total displacement of the bone segments.
METHODS: 3 Ilizarov configurations, with various wire pre-tensions and bolt-tightening torques, were tested 3 times with an MTS machine. For each wire, slippage and tensions were measured for the duration of the whole experiment.
RESULTS: A loss of wire pre-tension (up to 75%) due to slippage was found. Higher bolt-tightening torques significantly reduced the amount of pre-tension loss. Furthermore, a higher wire pre-tension reduced the maximal axial displacement of the bone fragment. There was, however, least cyclic axial micromotion when no wire pre-tension was present -- probably due to the lack of wire recoil.
INTERPRETATION: Wires in an experimental Ilizarov external fixator lose pre-tension within a limited period of time, and probably also in the clinical setting. This does not seem to lead to increased cyclic axial micromotions of the bone fragements. However, the question remains as to how excessive axial displacement of the bone fragment and other loading configurations will affect the process of fracture healing in a patient.

Entities:  

Mesh:

Year:  2007        PMID: 17966025     DOI: 10.1080/17453670710014356

Source DB:  PubMed          Journal:  Acta Orthop        ISSN: 1745-3674            Impact factor:   3.717


  6 in total

1.  Towards a solution of the wires' slippage problem of the Ilizarov external fixator.

Authors:  C Bairaktari; G Athanassiou; E Panagiotopoulos; D Deligianni
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-08-20

2.  Biomechanical Analysis of the Behaviour at the Metaphyseal-Diaphyseal Junction of Complex Tibial Plateau Fractures Using Two Circular Fixator Configurations.

Authors:  Gracielle S Cardoso; Renato Amorim; Francisco M Penha; Françoá J Horn; Carlos Rm Roesler; Jefferson Lb Marques
Journal:  Strategies Trauma Limb Reconstr       Date:  2020 Sep-Dec

3.  The use of twin-ring Ilizarov external fixator constructs: application and biomechanical proof-of principle with possible clinical indications.

Authors:  Theodoros B Grivas; Evangelos A Magnissalis
Journal:  J Orthop Surg Res       Date:  2011-08-11       Impact factor: 2.359

4.  The influence of a weight-bearing platform on the mechanical behavior of two Ilizarov ring fixators: tensioned wires vs. half-pins.

Authors:  Jan Gessmann; Mustafa Citak; Birger Jettkant; Thomas A Schildhauer; Dominik Seybold
Journal:  J Orthop Surg Res       Date:  2011-12-12       Impact factor: 2.359

5.  Measurement of wire deflection on loading may indicate union in Ilizarov constructs, an in vitro model.

Authors:  Beth Lineham; Todd Stewart; Paul Harwood
Journal:  Strategies Trauma Limb Reconstr       Date:  2018-02-02

6.  Measurement of Wire Deflection on Loading may Indicate Union in Ilizarov Constructs: A Pilot Study.

Authors:  Beth Lineham; Todd Stewart; John Ward; Paul Harwood
Journal:  Strategies Trauma Limb Reconstr       Date:  2021 Sep-Dec
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.