Comparison of electrical velocimetry and thermodilution techniques for the measurement of cardiac output.

N Zoremba1, J Bickenbach, B Krauss, R Rossaint, R Kuhlen, G Schälte.   

Abstract

AIM: To compare a new method of non-invasive determination of cardiac output based on electrical velocimetry (EV-CO) with invasive thermodilution methods.
METHODS: Fifty critically ill patients were enrolled into the study. EV-CO was compared with cardiac output measured by a pulmonary artery catheter (PA-CO) in one group (n= 25) and by a femoral artery catheter (PiCCO-CO) in a second group (n= 25), by simultaneous measurements. Standard electrocardiography electrodes were used for non-invasive measurements, and EV-CO was calculated using the Bernstein-Osypka equation. The invasive measurements of PA-CO and PiCCO-CO were made by the injection of iced 0.9% saline and the recording of thermodilution curves.
RESULTS: The precision values of EV-CO, PA-CO and PiCCO-CO measurements were +/- 0.46 [95% confidence interval (95% CI), +/- 0.06], +/- 0.57 (95% CI, +/- 0.09) and +/- 0.48 l/min (95% CI, +/- 0.08 l/min), respectively. The mean differences between EV-CO and PA-CO or PiCCO-CO were -0.05 +/- 0.71 and 0.22 +/- 0.78 l/min, respectively. The lower and upper limits of agreement for the comparison of EV-CO with PA-CO were -1.47 and 1.37 l/min (95% CI, +/- 0.25 l/min), respectively. In the comparison of EV-CO and PiCCO-CO, lower and upper limits of -1.34 and 1.78 l/min (95% CI, +/- 0.27 l/min) were found. The percentage errors between EV-CO and PA-CO or PiCCO-CO were 26.5% and 26.4%, respectively.
CONCLUSIONS: The values of cardiac output were statistically comparable between the groups. Therefore, electrical velocimetry is a suitable method to evaluate haemodynamic variables with clinically acceptable accuracy.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17944633     DOI: 10.1111/j.1399-6576.2007.01445.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  32 in total

1.  A comparative evaluation of electrical velocimetry and inert gas rebreathing for the non-invasive assessment of cardiac output.

Authors:  Frederik Trinkmann; Manuel Berger; Ursula Hoffmann; Martin Borggrefe; Jens J Kaden; Joachim Saur
Journal:  Clin Res Cardiol       Date:  2011-07-01       Impact factor: 5.460

2.  Accuracy and precision of non-invasive cardiac output monitoring by electrical cardiometry: a systematic review and meta-analysis.

Authors:  M Sanders; S Servaas; C Slagt
Journal:  J Clin Monit Comput       Date:  2019-06-07       Impact factor: 2.502

3.  Hemodynamic alterations recorded by electrical cardiometry during ligation of ductus arteriosus in preterm infants.

Authors:  Reyin Lien; Kai-Hsiang Hsu; Jaw-Ji Chu; Yu-Sheng Chang
Journal:  Eur J Pediatr       Date:  2014-10-15       Impact factor: 3.183

4.  Comparison of electrical velocimetry and cardiac magnetic resonance imaging for the non-invasive determination of cardiac output.

Authors:  Frederik Trinkmann; Manuel Berger; Christina Doesch; Theano Papavassiliu; Stefan O Schoenberg; Martin Borggrefe; Jens J Kaden; Joachim Saur
Journal:  J Clin Monit Comput       Date:  2015-06-27       Impact factor: 2.502

5.  Influence of patent ductus arteriosus and ventilators on electrical velocimetry for measuring cardiac output in very-low/low birth weight infants.

Authors:  T Torigoe; S Sato; Y Nagayama; T Sato; H Yamazaki
Journal:  J Perinatol       Date:  2015-01-29       Impact factor: 2.521

Review 6.  Reliable and developmentally appropriate study end points are needed to achieve drug development for treatment of pediatric pulmonary arterial hypertension.

Authors:  H Sun; N Stockbridge; R L Ariagno; D Murphy; R M Nelson; W Rodriguez
Journal:  J Perinatol       Date:  2016-07-14       Impact factor: 2.521

7.  Stroke volume and cardiac output evaluation by electrical cardiometry: accuracy and reference nomograms in hemodynamically stable preterm neonates.

Authors:  A Boet; G Jourdain; S Demontoux; D De Luca
Journal:  J Perinatol       Date:  2016-04-21       Impact factor: 2.521

8.  Hemodynamic reference for neonates of different age and weight: a pilot study with electrical cardiometry.

Authors:  K-H Hsu; T-W Wu; Y-C Wang; W-H Lim; C-C Lee; R Lien
Journal:  J Perinatol       Date:  2016-02-18       Impact factor: 2.521

9.  The Role of Biomarkers and Surrogate End Points in Drug Development for Neonatal Pulmonary Arterial Hypertension.

Authors:  Haihao Sun; Norman Stockbridge; Ronald L Ariagno; Dianne Murphy
Journal:  Neoreviews       Date:  2016-02-01

10.  Clinical validation of a real-time data processing system for cardiac output and arterial pressure measurement during intraoperative biventricular pacing optimization.

Authors:  Christopher K Johnson; Santos E Cabreriza; Rana L Sahar; Alexander Rusanov; Daniel Y Wang; Bin Cheng; Mira S Gendy; T Alexander Quinn; Henry Michael Spotnitz
Journal:  ASAIO J       Date:  2012 May-Jun       Impact factor: 2.872

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.