Preclinical assessment of the long-term endurance of cemented hip stems. Part 2: in-vitro and ex-vivo fatigue damage of the cement mantle.

L Cristofolini1, P Erani, P Savigni, B Bordini, M Viceconti.   

Abstract

Fatigue damage in the cement mantle surrounding hip stems has been studied in the past. However, so far no quantitative method has been validated for assessing ex-vivo damage and for predicting the in-vitro risk of cement fracture. This work presents a method for measuring cement damage; the cement mantle was sliced and sections were inspected with dye penetrants and an optical microscope. Cracks were counted, measured, and classified by type in each region of the cement mantle. Statistical indicators (in total and per unit volume of cement) were proposed that allow quantitative comparison. The method was first validated on two implant types with known clinical success rate, which were tested in vitro using a physiological loading profile (described in Part 1 of this work). The most relevant indicators were able to detect statistical differences between the two designs. Retrieved cement mantles (the same design as one of the in-vitro stems) from revision surgery were also processed with the same inspection method. Excellent qualitative and quantitative agreement was found between the in-vitro generated fatigue damage and the cracking pattern found in the ex-vivo retrieved cement mantles. This demonstrated the effectiveness of the cement inspection protocol and provided a further validation to the in-vitro testing method.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17937198     DOI: 10.1243/09544119JEIM184

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  4 in total

1.  Using 'subcement' to simulate the long-term fatigue response of cemented femoral stems in a cadaver model: could a novel preclinical screening test have caught the Exeter matt problem?

Authors:  A Race; M A Miller; K A Mann
Journal:  Proc Inst Mech Eng H       Date:  2010       Impact factor: 1.617

2.  Interface micromotion of uncemented femoral components from postmortem retrieved total hip replacements.

Authors:  Kenneth A Mann; Mark A Miller; Peter A Costa; Amos Race; Timothy H Izant
Journal:  J Arthroplasty       Date:  2011-07-01       Impact factor: 4.757

3.  The posterior approach reduces the risk of thin cement mantles with a straight femoral stem design.

Authors:  Gavin J Macpherson; Christian Hank; Michael Schneider; Morag Trayner; Robert Elton; Colin R Howie; Steffen J Breusch
Journal:  Acta Orthop       Date:  2010-06       Impact factor: 3.717

4.  A modified PMMA cement (Sub-cement) for accelerated fatigue testing of cemented implant constructs using cadaveric bone.

Authors:  Amos Race; Mark A Miller; Kenneth A Mann
Journal:  J Biomech       Date:  2008-09-05       Impact factor: 2.712

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.