Comparison of delivery characteristics from a combination metered-dose inhaler using the Andersen cascade impactor and the next generation pharmaceutical impactor.

Changning Guo1, Stacey R Gillespie, John Kauffman, William H Doub.   

Abstract

The purpose of this research was to compare two cascade impaction devices for the aerodynamic particle size assessment of a combination metered-dose inhaler (MDI) product, Combivent. Particle size analysis was performed using an Anderson Mark II cascade impactor (ACI) and a Next Generation Pharmaceutical Impactor (NGI), both fitted with a preseparator and either a 1 L glass chamber or USP throat, and operated at various flow rates. Particle size distributions (PSDs) and dose delivery profiles were assessed by means of the mass median aerodynamic diameter (MMAD), geometric standard deviation (GSD), fine particle fraction <5 micron aerodynamic diameter (FPF(<5 microm)), and induction port deposition fraction (IPF). Under their normal operating conditions, the ACI (28.3 L/min) and the NGI (30 L/min) yield similar PSDs and dose delivery profiles. However, this equivalent performance for the ACI and the NGI no longer exists at a higher flow rate of 60 L/min. Furthermore, changes in PSD results may also be obtained between different operators and/or when different induction port designs were employed. Thus, it is strongly recommended that special care be taken to eliminate variation in experimental parameters and/or selection of ancillary devices such as the preseparator, induction port or throat, to insure good repeatability and reproducibility when testing inhalation drugs.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 17932959     DOI: 10.1002/jps.21203

Source DB:  PubMed          Journal:  J Pharm Sci        ISSN: 0022-3549            Impact factor:   3.534


  3 in total

1.  The comparison of fluid dynamics parameters in an Andersen cascade impactor equipped with and without a preseparator.

Authors:  Janwit Dechraksa; Tan Suwandecha; Kittinan Maliwan; Teerapol Srichana
Journal:  AAPS PharmSciTech       Date:  2014-04-24       Impact factor: 3.246

2.  Influence of Formulation Factors on the Aerosol Performance of Suspension and Solution Metered Dose Inhalers: A Systematic Approach.

Authors:  Poonam Sheth; Dennis Sandell; Denise S Conti; Jay T Holt; Anthony J Hickey; Bhawana Saluja
Journal:  AAPS J       Date:  2017-06-07       Impact factor: 4.009

3.  Efficiency of Ipratropium Bromide and Albuterol Deposition in the Lung Delivered via a Soft Mist Inhaler or Chlorofluorocarbon Metered-Dose Inhaler.

Authors:  T R MacGregor; R ZuWallack; V Rubano; M A Castles; H Dewberry; M Ghafouri; C C Wood
Journal:  Clin Transl Sci       Date:  2016-03-06       Impact factor: 4.689

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.