The determination of hydrogen peroxide generated from cigarette smoke with an ultrasensitive and highly selective chemiluminescence method.

Yufei Hu1, Zhujun Zhang, Chunyan Yang.   

Abstract

Monitoring hydrogen peroxide (H(2)O(2)) in aqueous cigarette smoke condensate (CSC) is helpful for interpreting the relationship between cigarette smoke and oxidative stress, inflammation and disease. It is also significative for elucidating the pathogenic effects of CSC. In this paper, a novel flow-injection chemiluminescence (FI-CL) method was well established for determination of H(2)O(2) in the complex sample CSC which did not need pretreatment. The sensitive and selective method is based on the CL reaction of luminol with low concentration (10(-7) molL(-1)) and H(2)O(2) at low concentration level (<10(-8) molL(-1)) in an alkaline medium catalyzed by a complex K(5)[Cu(HIO(6))(2)] (DPC), which has proved no interference of other metal ions or horseradish peroxidase (HRP). The proposed method had been used to determine trace amount of H(2)O(2) with a limit of detection (3sigma) of 4.1 x 10(-11) molL(-1), which enables minimal amount of sample for analysis. A satisfactory result has been gained for the determination of H(2)O(2) in CSC sample by use of the proposed method. The concentration of H(2)O(2) in two reference cigarette (84 cm, Longfeng) smoke condensate have been determined at 4-6 micromolL(-1) level.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17904474     DOI: 10.1016/j.aca.2007.08.018

Source DB:  PubMed          Journal:  Anal Chim Acta        ISSN: 0003-2670            Impact factor:   6.558


  4 in total

1.  Study of enhanced chemiluminescence of diperiodatocuprate (III) on 1,10-phenanthroline/hydrogen peroxide/cetyltrimethylammonium bromide system.

Authors:  Jing Kang; Jieli Tang; Lu Han; Yihua Zhang
Journal:  J Fluoresc       Date:  2010-12-03       Impact factor: 2.217

2.  Ultra-small dispersed Cu x O nanoparticles on graphene fibers for miniaturized electrochemical sensor applications.

Authors:  Jinfeng Zeng; Xiaoteng Ding; Liwei Chen; Le Jiao; Yuze Wang; Christopher D Windle; Qing Han; Liangti Qu
Journal:  RSC Adv       Date:  2019-09-09       Impact factor: 4.036

3.  A novel luminol-based chemiluminescence method for the determination of amikacin sulfate in serum by using trivalent copper-periodate complex.

Authors:  Yu-Fei Hu; Gong-Ke Li; Zhu-Jun Zhang
Journal:  J Pharm Anal       Date:  2013-01-11

4.  Direct chemical vapor deposition of graphene on plasma-etched quartz glass combined with Pt nanoparticles as an independent transparent electrode for non-enzymatic sensing of hydrogen peroxide.

Authors:  Ning Li; Yawen Yuan; Jinglei Liu; Shifeng Hou
Journal:  RSC Adv       Date:  2020-05-28       Impact factor: 4.036

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.