Elena Vinogradov1, Huamei He, Angelo Lubag, James A Balschi, A Dean Sherry, Robert E Lenkinski.
Abstract
In this report, the On resonance PARamagnetic CHemical Exchange Effects (OPARACHEE) method was implemented in vivo using WALTZ-16* as a preparation pulse with a standard spin echo sequence to detect the accumulation and clearance of the TmDOTA-4AmC(-) in mouse kidney. The performance of the technique in vivo is described in terms of the magnitude of the contrast effect versus the bolus agent concentration and signal-to-noise ratio (SNR) levels. The lowest injected concentration of TmDOTA-4AmC(-), 200 microL of a 2-mM stock solution (corresponds to approximately 0.2 mM agent in plasma), reduced the total water signal in the kidney papilla by 45% 3 min after the a bolus injection. The results show that the OPARACHEE methodology employing low-amplitude RF trains can detect paramagnetic exchanging agents in vivo.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17899603 DOI: 10.1002/mrm.21393
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668