Accuracy of cerebral monitoring in detecting cerebral ischemia during carotid endarterectomy: a comparison of transcranial Doppler sonography, near-infrared spectroscopy, stump pressure, and somatosensory evoked potentials.

Stefan Moritz1, Piotr Kasprzak, Matthias Arlt, Kai Taeger, Christoph Metz.   

Abstract

BACKGROUND: This study compares the accuracy of cerebral monitoring systems in detecting cerebral ischemia during carotid endarterectomy.
METHODS: The authors compared transcranial Doppler sonography (TCD), near-infrared spectroscopy (NIRS), stump pressure (SP) measurement, and somatosensory evoked potentials (SEP) in 48 patients undergoing carotid surgery during regional anesthesia. Cerebral ischemia was assumed when neurologic deterioration occurred. During clamping, the minimum mean middle cerebral artery velocity (TCD(min)), its percentage change (TCD%), the minimum regional saturation of oxygen (NIRS(min)), its percentage change (NIRS%), the mean SP, and the changes of SEP amplitude were recorded. To analyze the corresponding sensitivity and specificity of each parameter, the authors performed receiver operating characteristic analysis.
RESULTS: Neurologic deterioration occurred in 12 patients. SP and NIRS were successfully performed in all patients. TCD monitoring was not possible in 10 (21%); SEP was not possible in 2 patients (4%). All parameters provided the ability to distinguish between ischemic and nonischemic patients. TCD% and NIRS% showed significantly better discrimination than TCD(min) and NIRS(min) (P < 0.05). The highest area under the curve (AUC) was found for TCD% (AUC = 0.973), but there was no significant difference compared with NIRS% (AUC = 0.905) and SP (AUC = 0.925). The lowest AUC was found for SEP (AUC = 0.749), which was significantly lower than that for TCD%, NIRS%, and SP.
CONCLUSIONS: TCD%, NIRS%, and SP measurement provide similar accuracy for the detection of cerebral ischemia during carotid surgery. Lower accuracy was found for SEP monitoring. Because of the high rate of technical difficulties (21%), TCD monitoring was the least practical of the investigated monitoring devices.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17893451     DOI: 10.1097/01.anes.0000281894.69422.ff

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  53 in total

1.  Usefulness of near-infrared spectroscopy in thrombectomy monitoring.

Authors:  Thomas Ritzenthaler; Tae-Hee Cho; David Luis; Yves Berthezene; Norbert Nighoghossian
Journal:  J Clin Monit Comput       Date:  2014-11-04       Impact factor: 2.502

2.  The usefulness of near-infrared spectroscopy in the anesthetic management of endovascular aortic aneurysm repair.

Authors:  Masahiro M Wakimoto; Mamoru Kadosaki; Hirofumi Nagata; Kenji S Suzuki
Journal:  J Anesth       Date:  2012-06-26       Impact factor: 2.078

3.  Suprasternal approach for impending tracheo-innominate artery fistula.

Authors:  Yoshifumi Fujimoto; Keiichi Hirose; Noritaka Ota; Masaya Murata; Yujiro Ide; Yuko Tosaka; Maiko Tachi; Kisaburo Sakamoto
Journal:  Gen Thorac Cardiovasc Surg       Date:  2010-09-22

4.  Effect of age on intraoperative cerebrovascular autoregulation and near-infrared spectroscopy-derived cerebral oxygenation.

Authors:  C S Burkhart; A Rossi; S Dell-Kuster; M Gamberini; A Möckli; M Siegemund; M Czosnyka; S P Strebel; L A Steiner
Journal:  Br J Anaesth       Date:  2011-08-10       Impact factor: 9.166

5.  Application of full field optical studies for pulsatile flow in a carotid artery phantom.

Authors:  M Nemati; G B Loozen; N van der Wekken; G van de Belt; H P Urbach; N Bhattacharya; S Kenjeres
Journal:  Biomed Opt Express       Date:  2015-09-21       Impact factor: 3.732

6.  Different ventilation techniques and hemodynamic optimization to maintain regional cerebral oxygen saturation (rScO2) during laparoscopic bariatric surgery: a prospective randomized interventional study.

Authors:  Osama M Asaad
Journal:  J Anesth       Date:  2018-04-06       Impact factor: 2.078

7.  Evaluation of near-infrared spectroscopy under apnea-dependent hypoxia in humans.

Authors:  Lars Eichhorn; Felix Erdfelder; Florian Kessler; Jonas Doerner; Marcus O Thudium; Rainer Meyer; Richard K Ellerkmann
Journal:  J Clin Monit Comput       Date:  2015-02-04       Impact factor: 2.502

8.  Limit of intraoperative near-infrared spectroscopy monitoring during endovascular thrombectomy in acute ischemic stroke.

Authors:  Ryo Hiramatsu; Motomasa Furuse; Ryokichi Yagi; Hiroyuki Ohnishi; Naokado Ikeda; Naosuke Nonoguchi; Shinji Kawabata; Shigeru Miyachi; Toshihiko Kuroiwa
Journal:  Interv Neuroradiol       Date:  2017-11-29       Impact factor: 1.610

Review 9.  [Cerebral oximetry: clinical importance for cardiac surgery patients].

Authors:  J Schön; H Paarmann; M Heringlake
Journal:  Anaesthesist       Date:  2012-09-12       Impact factor: 1.041

Review 10.  Cerebral and tissue oximetry.

Authors:  Jochen Steppan; Charles W Hogue
Journal:  Best Pract Res Clin Anaesthesiol       Date:  2014-09-28
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.