Evaluation of the SCHILLER BR-102 plus noninvasive ambulatory blood pressure monitor according to the International Protocol introduced by the Working Group on Blood Pressure Monitoring of the European Society of Hypertension.

Stefan V Denchev1, Iana I Simova, Mikhail G Matveev.   

Abstract

OBJECTIVE: To evaluate the SCHILLER BR-102 plus (Schiller AG, Baar, Switzerland) noninvasive ambulatory blood pressure recorder according to the International Protocol for validation of blood pressure measuring devices in adults introduced by the Working Group on Blood Pressure Monitoring of the European Society of Hypertension.
METHOD: One SCHILLER BR-102 plus blood pressure recorder was tested by a validation team, consisting of three persons: two observers (nurses) and a doctor, acting as supervisor and 'expert'. The European Society of Hypertension International Protocol comprises two phases. Fifteen participants were recruited for the first phase and, following a successful test, a further 18 persons (giving a total of 33) were recruited additionally. For phase 1, five of the 15 participants had systolic blood pressure and diastolic blood pressure in each of the ranges: low, medium and high. For phase 2, 11 of the 33 participants (including the first 15 participants) had systolic blood pressure and diastolic blood pressure in each of the ranges. The mercury standard for validation was preferred over the optional Sphygmocorder. Because the SCHILLER BR-102 plus has the oscillometric method as backup to the basic auscultatory measurement, both systems of measurement were subjected to individual validations.
RESULTS: Results obtained show that the SCHILLER BR-102 plus noninvasive ambulatory blood pressure recorder meets all the requirements specified in the International Protocol for both oscillometric and auscultatory methods.
CONCLUSION: On the basis of these results, the SCHILLER BR-102 plus can be recommended for ambulatory blood pressure measurement in clinical practice using both auscultatory and oscillometric modes.

Entities:  

Mesh:

Year:  2007        PMID: 17890972     DOI: 10.1097/MBP.0b013e32813fa39e

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  4 in total

1.  Blood pressure profiles 5 to 10 years after transplant in pediatric solid organ recipients.

Authors:  Juuso Tainio; Erik Qvist; Jenni Miettinen; Tuula Hölttä; Mikko Pakarinen; Timo Jahnukainen; Hannu Jalanko
Journal:  J Clin Hypertens (Greenwich)       Date:  2015-01-05       Impact factor: 3.738

2.  Validation of the custo screen 400 ambulatory blood pressure-monitoring device according to the European Society of Hypertension International Protocol revision 2010.

Authors:  Peter Bramlage; Cornelia Deutsch; Ralf Krüger; Andreas Wolf; Peter Müller; Thomas Zwingers; Beate Beime; Thomas Mengden
Journal:  Vasc Health Risk Manag       Date:  2014-05-13

3.  Twenty-Four-Hour Ambulatory Blood Pressure Monitoring.

Authors:  Camilo Pena-Hernandez; Kenneth Nugent; Meryem Tuncel
Journal:  J Prim Care Community Health       Date:  2020 Jan-Dec

4.  Whole blood omega-3 fatty acid concentrations are inversely associated with blood pressure in young, healthy adults.

Authors:  Mark G Filipovic; Stefanie Aeschbacher; Martin F Reiner; Simona Stivala; Sara Gobbato; Nicole Bonetti; Martin Risch; Lorenz Risch; Giovanni G Camici; Thomas F Luescher; Clemens von Schacky; David Conen; Juerg H Beer
Journal:  J Hypertens       Date:  2018-07       Impact factor: 4.844

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.