Rebecca L Rich1, Michelle J Cannon, Jerry Jenkins, Prabhakar Pandian, Shankar Sundaram, Rachelle Magyar, Jennifer Brockman, Jeremy Lambert, David G Myszka.
Abstract
Surface plasmon resonance imaging systems, such as Flexchip from Biacore, are capable of monitoring hundreds of reaction spots simultaneously within a single flow cell. Interpreting the binding kinetics in a large-format flow cell presents a number of potential challenges, including accounting for mass transport effects and spot-to-spot sample depletion. We employed a combination of computer simulations and experimentation to characterize these effects across the spotted array and established that a simple two-compartment model may be used to accurately extract intrinsic rate constants from the array under mass transport-limited conditions. Using antibody systems, we demonstrate that the spot-to-spot variability in the binding kinetics was <9%. We also illustrate the advantage of globally fitting binding data from multiple spots within an array for a system that is mass transport limited.Mesh:
Substances:
Year: 2007 PMID: 17889820 DOI: 10.1016/j.ab.2007.08.017