Literature DB >> 1788845

Leaflet embolisation from Duromedics valves: a report of two cases.

N Kumar1, S Balasundaram, M Rickard, Z al Halees, C M Duran.   

Abstract

Embolization of parts of mechanical valves has been reported since the inception of prosthetic valve implantation. We report here two cases of embolization of one hemileaflet of a Duromedic bileaflet prosthesis in the mitral position due to a pivot fracture. Both presented with moderately severe mitral regurgitation and pulmonary edema and were successfully managed by replacement of the malfunctioning prostheses. The embolised disc was located in the left common iliac artery by abdominal ultrasound and removed by an inguinal, retroperitoneal approach with low morbidity. Both patients left hospital and are doing well to-date.

Entities:  

Mesh:

Year:  1991        PMID: 1788845     DOI: 10.1055/s-2007-1020005

Source DB:  PubMed          Journal:  Thorac Cardiovasc Surg        ISSN: 0171-6425            Impact factor:   1.827


  3 in total

1.  Prosthetic mitral valve leaflet escape.

Authors:  Darae Kim; Sin Sang Hun; In-Jeong Cho; Chi-Young Shim; Jong-Won Ha; Namsik Chung; Hyun Chul Ju; Jang Won Sohn; Geu-Ru Hong
Journal:  J Cardiovasc Ultrasound       Date:  2013-06-26

2.  Long-term event-free survival with an embolised prosthetic valve leaflet in the thoracic aorta.

Authors:  Mohan P Devbhandari; Edwin B C Woo; Timothy L Hooper
Journal:  J Cardiothorac Surg       Date:  2008-05-30       Impact factor: 1.637

3.  Spontaneous leaflet fracture resulting in embolization from mechanical valve prostheses.

Authors:  Gijs G J van Steenbergen; Queeny H Y Tsang; Stefan M van der Heide; Michel W A Verkroost; Wilson W L Li; Wim J Morshuis
Journal:  J Card Surg       Date:  2018-12-30       Impact factor: 1.620

  3 in total