Locking plates increase the strength of dynamic hip screws.

Dylan P A Jewell1, Sabina Gheduzzi, Mark S Mitchell, Anthony W Miles.   

Abstract

INTRODUCTION: Failure of a dynamic hip screw (DHS) fixation leads to decreased mobility of the patient and frequently to a decrease in general health. The most common mode of failure of a DHS is cut out of the lag screw from the femoral head. The second most common mode of failure is lift-off of the plate from the femur. The aim of this laboratory-based experimental study was to determine whether a DHS secured to an osteoporotic femur with a locking screw plate would provide a stronger construct than the standard DHS plate.
METHOD: The standard DHS design was compared to a DHS with fixed angle locking screws holding the DHS plate to the femur. Standard dynamic compression plates (DCP) and locking compression plates (LCP) were attached to synthetic, osteoporotic bone. A load was applied to replicate the forces occurring following the fixation of unstable, intertrochanteric hip fractures. A bracket on the proximal end of the plate replicated the lag screw in the femoral head. The constructs were cyclically loaded by a screw-driven material-testing machine and the number of cycles before failure occurred was determined.
RESULTS: The mean number of cycles to failure for the locking plate construct was 2.6 times greater than for the standard screw construct (285 versus 108 cycles, respectively p=0.016).
CONCLUSION: A dynamic hip screw with fixed angle locking screws would reduce the risk of DHS failure. A locking screw DHS would be particularly useful in patients with osteoporotic bone, and in patients with less stable fracture configurations.

Entities:  

Mesh:

Year:  2007        PMID: 17880976     DOI: 10.1016/j.injury.2007.05.018

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  4 in total

1.  Mechanical evaluation of fourth-generation composite femur hybrid locking plate constructs.

Authors:  Tarun Goswami; Vinit Patel; David J Dalstrom; Michael J Prayson
Journal:  J Mater Sci Mater Med       Date:  2011-07-17       Impact factor: 3.896

2.  A comparative Study of Novel Extramedullary Fixation and Dynamic Hip Screw in the Fixation of Intertrochanteric Fracture: A Finite-Element Analysis.

Authors:  Kai Ding; Yanbin Zhu; Haicheng Wang; Yonglong Li; Weijie Yang; Xiaodong Cheng; Yingze Zhang; Wei Chen; Qi Zhang
Journal:  Front Surg       Date:  2022-05-25

3.  Tips and Tricks in Extracapsular Hip Fracture Fixation.

Authors:  Ankit Desai; Josh Jacob; Alex Trompeter
Journal:  Open Orthop J       Date:  2017-10-31

4.  Radiological and functional outcome in unstable, osteoporotic trochanteric fractures stabilized with dynamic helical hip system.

Authors:  Ram Chander Siwach; Rajesh Rohilla; Roop Singh; Rohit Singla; Sukhbir Singh Sangwan; Paritosh Gogna
Journal:  Strategies Trauma Limb Reconstr       Date:  2013-07-28