Literature DB >> 1787939

Retinal tolerance of an implantable light source for use during vitrectomy surgery.

J F Vander1, R C Eagle, G C Brown, V Arbizo, E P Shakin, R Reber.   

Abstract

To assess for the possible development of a phototoxic lesion, we exposed five Dutch belted rabbit retinas to constant illumination with an implantable "chandelier" light source at maximal intensity for 1 hour. No clinically apparent lesion developed in any of the experimentally treated eyes by 48 to 72 hours posttreatment. Light and electron microscopic examination failed to reveal any of the characteristic changes of retinal phototoxicity. A control eye treated with a 20-minute exposure to a standard endoiluminator probe at maximal intensity held 2 mm from the retinal surface demonstrated typical light and electron microscopic phototoxic abnormalities, whereas a retina similarly treated with a 5-minute exposure was normal. We have confirmed that a standard endoilluminator probe can create a phototoxic lesion in the experimental setting. Furthermore, no phototoxicity from the sustained retinal exposure to a fixed, implantable light source for a clinically relevant duration could be demonstrated.

Entities:  

Mesh:

Year:  1991        PMID: 1787939

Source DB:  PubMed          Journal:  Ophthalmic Surg        ISSN: 0022-023X


  4 in total

1.  Evaluation of the safety of xenon/bandpass light in vitrectomy using the A2E-laden RPE model.

Authors:  Yasuo Yanagi; Aya Iriyama; Woo-Dong Jang; Kazuaki Kadonosono
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-10-17       Impact factor: 3.117

Review 2.  [Exposure to light during vitreoretinal surgery. II: Characteristics of endoilluminators].

Authors:  T Ach; A E Höh; R Amberger; S Dithmar
Journal:  Ophthalmologe       Date:  2008-10       Impact factor: 1.059

3.  Involvement of illumination in indocyanine green toxicity after its washout in the ex vivo rat retina.

Authors:  Kazuhiro Tokuda; Charles F Zorumski; Yukitoshi Izumi
Journal:  Retina       Date:  2009-03       Impact factor: 4.256

4.  Management of posteriorly dislocated capsular tension ring with vitrectomy using a "chandelier" illumination system.

Authors:  Amar Agarwal; Rupal H Trivedi; Soosan Jacob; Mandeep Lamba; Athiya Agarwal; Sunita Agarwal
Journal:  Clin Ophthalmol       Date:  2007-09