Literature DB >> 17768084

Evaluation of the analytical performance of the new Abbott RealTime RT-PCRs for the quantitative detection of HCV and HIV-1 RNA.

M Schutten1, E Fries, C Burghoorn-Maas, H G M Niesters.   

Abstract

BACKGROUND: Despite FDA approval and CE marking of commercial tests, manufacturer independent testing of technical aspects is important.
OBJECTIVES: To evaluate the analytical performance of the new Abbott RealTime HCV and HIV-1 viral load tests. STUDY
DESIGN: Sensitivity, specificity and inter-/intra-assay variation were investigated. The HCV and HIV-1 assays were compared with Siemens bDNA 3.0 and Roche Cobas Monitor 2.0, respectively, on diagnostic samples.
RESULTS: Lower isolation volumes on the M1000 gave minor but statistically significant lower quantitative values. Minor differences were observed in the lower limit of detection relative to the specification given by the manufacturer. Inter-/intra-assay coefficients of variations ranged from 0.31 to 4.75 between 5.0 x 10(4) and 5.0 x 10(2) copies/mL. Both the HCV and HIV-1 Abbott RealTime tests did not show a geno-/sub-type dependent under-quantification on WHO reference panels, quality control panels or clinical specimens. The Abbott RealTime HIV-1 viral load assay detected subtype O whereas several other systems failed to detect this subtype.
CONCLUSION: The technical aspects of the HCV and HIV-1 RealTime viral load assays on the M2000 system make it attractive for use in routine diagnostic settings.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17768084     DOI: 10.1016/j.jcv.2007.07.013

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


  12 in total

1.  European Multicenter Study on Analytical Performance of Veris HIV-1 Assay.

Authors:  Patrick Braun; Rafael Delgado; Monica Drago; Diana Fanti; Hervé Fleury; Jörg Hofmann; Jacques Izopet; Ulrich Kalus; Alessandra Lombardi; Maria Angeles Marcos; Davide Mileto; Karine Sauné; Siobhan O'Shea; Alfredo Pérez-Rivilla; John Ramble; Pascale Trimoulet; Jordi Vila; Duncan Whittaker; Alain Artus; Daniel W Rhodes
Journal:  J Clin Microbiol       Date:  2017-04-19       Impact factor: 5.948

2.  Performance of the Abbott RealTime and Roche Cobas TaqMan hepatitis C virus (HCV) assays for quantification of HCV genotypes.

Authors:  Heidi LaRue; Lisa Rigali; Joan-Miquel Balada-Llasat; Preeti Pancholi
Journal:  J Clin Microbiol       Date:  2012-02-29       Impact factor: 5.948

3.  Quantification of viral DNA and liver enzymes in plasma improves early diagnosis and management of herpes simplex virus hepatitis.

Authors:  M F C Beersma; G M G M Verjans; H J Metselaar; A D M E Osterhaus; W R Berrington; G J van Doornum
Journal:  J Viral Hepat       Date:  2010-08-12       Impact factor: 3.728

4.  Performance evaluation of the new Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 test version 2.0 for quantification of human immunodeficiency virus type 1 RNA.

Authors:  Suzan Pas; John W A Rossen; Daniel Schoener; Diana Thamke; Annika Pettersson; Reiner Babiel; Martin Schutten
Journal:  J Clin Microbiol       Date:  2010-02-17       Impact factor: 5.948

5.  Abbott RealTime hepatitis C virus (HCV) and Roche Cobas AmpliPrep/Cobas TaqMan HCV assays for prediction of sustained virological response to pegylated interferon and ribavirin in chronic hepatitis C patients.

Authors:  Kentaro Matsuura; Yasuhito Tanaka; Izumi Hasegawa; Tomoyoshi Ohno; Hiroshi Tokuda; Fuat Kurbanov; Fuminaka Sugauchi; Shunsuke Nojiri; Takashi Joh; Masashi Mizokami
Journal:  J Clin Microbiol       Date:  2008-12-17       Impact factor: 5.948

6.  Evaluation of the Abbott m2000 RealTime human immunodeficiency virus type 1 (HIV-1) assay for HIV load monitoring in South Africa compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 assays.

Authors:  Lesley E Scott; Lara D Noble; Jackie Moloi; Linda Erasmus; Willem D F Venter; Wendy Stevens
Journal:  J Clin Microbiol       Date:  2009-05-06       Impact factor: 5.948

7.  Differences between two real-time PCR-based hepatitis C virus (HCV) assays (RealTime HCV and Cobas AmpliPrep/Cobas TaqMan) and one signal amplification assay (Versant HCV RNA 3.0) for RNA detection and quantification.

Authors:  Johannes Vermehren; Annika Kau; Barbara C Gärtner; Reinhild Göbel; Stefan Zeuzem; Christoph Sarrazin
Journal:  J Clin Microbiol       Date:  2008-09-17       Impact factor: 5.948

8.  Evaluation of an automated nucleic acid extractor for hepatitis C virus load quantification.

Authors:  Elisa Martró; Nerea García-Sierra; Victoria González; Verónica Saludes; Lurdes Matas; Vicenç Ausina
Journal:  J Clin Microbiol       Date:  2009-01-07       Impact factor: 5.948

9.  Performance of the Abbott real-time PCR assay using m2000sp and m2000rt for hepatitis C virus RNA quantification.

Authors:  Stéphane Chevaliez; Magali Bouvier-Alias; Jean-Michel Pawlotsky
Journal:  J Clin Microbiol       Date:  2009-04-15       Impact factor: 5.948

10.  Real-Time PCR Assays for the Quantification of HCV RNA: Concordance, Discrepancies and Implications for Response Guided Therapy.

Authors:  Robert Strassl; Karoline Rutter; Albert Friedrich Stättermayer; Sandra Beinhardt; Michael Kammer; Harald Hofer; Peter Ferenci; Theresia Popow-Kraupp
Journal:  PLoS One       Date:  2015-08-14       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.