Quality assurance of brachytherapy afterloaders using a multi-slit phantom.

Peter Kohr1, Frank-André Siebert.   

Abstract

Brachytherapy is a very successful treatment option for many different tumors. Consequently the quality assurance (QA) of afterloaders is of special interest. For many quality checks verification films are essential. Because currently in many clinics the film development units are replaced by computed radiography systems, the question arises of how to perform QA for afterloaders without conventional radiography films. In this note the use of a new designed multi-slit phantom demonstrates the possibility of performing QA checks with high sensitivity computed radiography systems. Because the presented phantom allows a method for imaging the radiation source it might be useful in the development of further QA techniques.

Entities:  

Mesh:

Year:  2007        PMID: 17762074     DOI: 10.1088/0031-9155/52/17/N03

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  5 in total

1.  Dose distribution verification for GYN brachytherapy using EBT Gafchromic film and TG-43 calculation.

Authors:  Somayeh Gholami; Hamid Reza Mirzaei; Ali Jabbary Arfaee; Ramin Jaberi; Hassan Ali Nedaie; Seied Rabi Mahdavi; Eftekhar Rajab Bolookat; Ali S Meigooni
Journal:  Rep Pract Oncol Radiother       Date:  2016-07-18

2.  Comparison of catheter reconstruction techniques for the lunar ovoid channels of the VeneziaTM applicator.

Authors:  Jon Hansen; David Dunkerley; Kristin Bradley; Jessica Miller; Jessie Huang
Journal:  J Contemp Brachytherapy       Date:  2020-08-21

3.  HDR and PDR 192Ir source activity control procedures, as the part of the quality assurance system at Brachytherapy Department of Greater Poland Cancer Centre.

Authors:  Grzegorz Zwierzchowski; Barbara Błasiak; Patrycja Stefaniak; Grzegorz Bielęda
Journal:  J Contemp Brachytherapy       Date:  2009-10-08

Review 4.  Electromagnetic tracking for treatment verification in interstitial brachytherapy.

Authors:  Christoph Bert; Markus Kellermeier; Kari Tanderup
Journal:  J Contemp Brachytherapy       Date:  2016-11-02

Review 5.  Review of clinical brachytherapy uncertainties: analysis guidelines of GEC-ESTRO and the AAPM.

Authors:  Christian Kirisits; Mark J Rivard; Dimos Baltas; Facundo Ballester; Marisol De Brabandere; Rob van der Laarse; Yury Niatsetski; Panagiotis Papagiannis; Taran Paulsen Hellebust; Jose Perez-Calatayud; Kari Tanderup; Jack L M Venselaar; Frank-André Siebert
Journal:  Radiother Oncol       Date:  2013-11-30       Impact factor: 6.280

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.