| Literature DB >> 17720423 |
Yasuko Tomizawa1, Takao Hanawa.
Abstract
Titanium and its alloys are considered reliable materials commonly used in heart surgeries. A patient had severe inflammation at the median sternotomy wound develop after open heart surgery. Sternal wires were removed and were found to be seriously corroded on most of the surface. Energy dispersive spectroscopy identified titanium as the only elemental component of the explanted wire with grade 2 purity, and the polarization curve showed satisfactory corrosion resistance similar to commercial pure titanium. Prolonged inflammation might have caused extensive corrosion of titanium in the present case.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17720423 DOI: 10.1016/j.athoracsur.2007.04.049
Source DB: PubMed Journal: Ann Thorac Surg ISSN: 0003-4975 Impact factor: 4.330