Measuring the efficiency of an artificial lung: 1. Carbon dioxide transfer.

D A Pybus1, M Lyon, J Hamilton, M Henderson.   

Abstract

The performance of a hollow fibre artificial lung ('Capiox E') was analysed by measurement of the 'parallel deadspace' of the device under varying conditions in 21 patients. The efficiency with which carbon dioxide was exchanged was determined by the time available for equilibration between the blood and gas phases. When this equilibration coefficient was less than 12 seconds per litre of blood flow per litre of gas flow, there was a marked reduction in the efficiency of gas exchange. Under certain conditions, the 'counter-current' design of the device apparently permitted the clearance of carbon dioxide at a partial pressure greater than that which was found in the mixed venous blood. This anomalous behaviour may represent in vivo confirmation of the Haldane effect.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1767913     DOI: 10.1177/0310057X9101900320

Source DB:  PubMed          Journal:  Anaesth Intensive Care        ISSN: 0310-057X            Impact factor:   1.669


  1 in total

1.  "Orpheus" cardiopulmonary bypass simulation system.

Authors:  Richard W Morris; David A Pybus
Journal:  J Extra Corpor Technol       Date:  2007-12