Katri Roppola1, Toivo Kuokkanen, Jaakko Rämö, Hanna Prokkola, Eeva Heiska.
Abstract
The aim of this research was to study the suitability of a manometric respirometric test (OECD 301F) for the determination of biological oxygen demand (BOD) in domestic wastewater. This was done by comparing four different BOD tests (a manometric respirometric test, oxygen sensor determination, iodometric titration, and UV absorption measurement) and by evaluating the BOD(7) value after a short-term measurement applying the manometric respirometric test. Measurements were done in three different laboratories. The respirometric test was performed well on municipal sewage, and the precision of results was good. There were minor deviations between BOD(7) values as determined by the different methods, but the deviations were not essentially big enough to have a practical impact. Evaluation of the BOD(7) value after two or three days' incubation by the respirometric method was shown to work well; average deviations between measured and evaluated values after three days' incubation were 5-18%. Hence, the manometric respirometric test proved to be a rapid and suitable tool for process control purposes.Entities:
Year: 2007 PMID: 17671608 PMCID: PMC1920587 DOI: 10.1155/2007/39761
Source DB: PubMed Journal: J Autom Methods Manag Chem ISSN: 1463-9246
The BOD7 results at 20°C determined by respirometric test, iodometric titration, oxygen sensor, and UV absorption test for wastewaters from Taskila's WTP, Oulu Waterworks.
| Sample | Respirometric test [mg/L] | Iodometric test [mg/L] | Oxygen sensor [mg/L] | UV absorption | |
| Preclarified | 65 | 56 | — | — | |
| Biol.treated | 4.5 | 4.8 | — | 5.7 | |
| Outflowing | 5.6 | 5.6 | 5.0 | — | |
| Preclarified | 106 | 108 | — | — | |
| Biol.treated | 4.0 | 6.7 | — | 4.9 | |
| Outflowing | 4.2 | 5.8 | 4.0 | — | |
| Biol.treated | 6.2 | 4.5 | 5.0 | 10.6 | |
| Outflowing | 7.0 | 6.0 | 8.0 | — | |
| Preclarified | 86 | 66 | — | — | |
| Biol.treated | 5.3 | 6.0 | 4.0 | 2.1 | |
| Outflowing | 6.0 | 7.3 | 6.0 | — | |
(a) Results of the UV measurements (correlated to BOD7 value) are the mean values of one-day measurement.
Figure 1BOD results of duplication tests for pre-clarified, biologically treated and outflowing wastewaters from Oulu Waterworks, Taskila WTP at 20°C.
Evaluation results of wastewater examined in this research (equation number 1: y = ax + b; equation number 2: ; equation number 3: y = [(a + cx)/(1 + bx)]).
| Sample | Equation number | Evaluation of BOD7 after 2 days' incubation [mg/L] | Evaluation of BOD7 after 3 days' incubation [mg/L] | Evaluation of BOD7 after 4 days' incubation [mg/L] | BOD7 value [mg/L] | |
| Preclarified | 3 | 67 | 64 | 64 | 65 | |
| Biol.treated | 1 | 4.5 | 4.5 | 4.1 | 4.5 | |
| Biol.treated | 2 | 4.3 | 4.1 | 4.4 | 4.5 | |
| Outflowing | 1 | 6.2 | 5.8 | 5.7 | 5.6 | |
| Outflowing | 2 | 5.2 | 5.6 | 5.3 | 5.6 | |
| Preclarified | 3 | 98 | 106 | 104 | 108 | |
| Biol.treated | 1 | 3.7 | 4.1 | 4.0 | 4.0 | |
| Biol.treated | 2 | — | 3.3 | 3.4 | 4.0 | |
| Outflowing | 1 | 3.8 | 4.8 | 4.7 | 4.2 | |
| Outflowing | 2 | 2.2 | 4.4 | 4.6 | 4.2 | |
| Preclarified | 3 | 92 | 93 | 93 | 100 | |
| Biol.treated | 1 | 8.3 | 8.0 | 7.2 | 6.2 | |
| Biol.treated | 2 | 8.3 | 7.1 | 6.3 | 6.2 | |
| Outflowing | 1 | 9.7 | 9.2 | 8.8 | 7.0 | |
| Outflowing | 2 | 8.4 | 8.4 | 8.0 | 7.0 | |
| Preclarified | 3 | 113 | 94 | 88 | 86 | |
| Biol.treated | 2 | 7.2 | 6.0 | 5.6 | 5.3 | |
| Outflowing | 2 | 8.1 | 6.0 | 6.1 | 5.9 | |