Literature DB >> 17667236

Concentration of heparin-locking solution and risk of central venous hemodialysis catheter malfunction.

Chandra M Thomas1, Jianguo Zhang, Teik How Lim, Nairne Scott-Douglas, Ronald B Hons, Brenda R Hemmelgarn.   

Abstract

Heparin is used as an interdialytic locking solution for hemodialysis (HD) central venous catheters (CVCs). The purpose of this study was to compare effectiveness of two heparin concentrations (10,000 and 1,000 U/mL) in preventing catheter malfunction. We compared two time periods: a 6-month period with heparin 10,000 U/mL and a 3-month period with heparin 1,000 U/mL. Adults on HD using a CVC (tunneled or untunneled) in Calgary, Alberta, were included. The primary outcome was catheter malfunction. A total of 139 and 134 patients in the heparin 10,000 and 1,000 U/mL periods, respectively, were included. The crude rate of catheter malfunction, per 1,000 HD sessions, was similar for heparin 10,000 (7.6; 95% CI, 5.3 to 10.8) and 1,000 (6.7; 95% CI, 4.3 to 10.3) U/mL periods, respectively (p = 0.76). After adjusting for CVC characteristics and use of recombinant tissue plasminogen activator (rt-PA), there was no association between heparin concentration and CVC malfunction (hazard ratio, 0.77; 95% CI, 0.37 to 1.61). In conclusion, the use of a lower concentration of heparin was not associated with an increased risk of catheter malfunction but may be associated with greater rt-PA use. The association between heparin concentration and rt-PA use requires further study.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17667236     DOI: 10.1097/MAT.0b013e3180619519

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  6 in total

1.  Does the heparin lock concentration affect hemodialysis catheter patency?

Authors:  D Maya Ivan; Tamorie Smith; Michael Allon
Journal:  Clin J Am Soc Nephrol       Date:  2010-05-24       Impact factor: 8.237

Review 2.  Anticoagulants and antiplatelet agents for preventing central venous haemodialysis catheter malfunction in patients with end-stage kidney disease.

Authors:  Ying Wang; Jessica N Ivany; Vlado Perkovic; Martin P Gallagher; Mark Woodward; Meg J Jardine
Journal:  Cochrane Database Syst Rev       Date:  2016-04-04

3.  Minimizing hemodialysis catheter dysfunction: an ounce of prevention.

Authors:  Timmy Lee; Charmaine Lok; Miguel Vazquez; Louise Moist; Ivan Maya; Michele Mokrzycki
Journal:  Int J Nephrol       Date:  2012-02-19

4.  Risk factors associated with hemodialysis central venous catheter malfunction; a retrospective analysis of a randomized controlled trial.

Authors:  David R Ward; Louise M Moist; Jennifer M MacRae; Nairne Scott-Douglas; Jianguo Zhang; Marcello Tonelli; Charmaine E Lok; Steven D Soroka; Brenda R Hemmelgarn
Journal:  Can J Kidney Health Dis       Date:  2014-07-08

Review 5.  Reevaluation of lock solutions for Central venous catheters in hemodialysis: a narrative review.

Authors:  Yiqin Wang; Xuefeng Sun
Journal:  Ren Fail       Date:  2022-12       Impact factor: 3.222

6.  Sodium bicarbonate catheter lock solution reduces hemodialysis catheter loss due to catheter-related thrombosis and blood stream infection: an open-label clinical trial.

Authors:  Adel S El-Hennawy; Elena Frolova; Wesley A Romney
Journal:  Nephrol Dial Transplant       Date:  2019-10-01       Impact factor: 5.992