Literature DB >> 17665787

Vascular risk associated with meniscal repair using Rapidloc versus FasT-Fix: comparison of two all-inside meniscal devices.

Steven B Cohen1, Lucy Boyd, Mark D Miller.   

Abstract

Eight fresh-frozen cadaver legs (average age 72 years) were used to evaluate the risk to the popliteal artery related to the use of two all-inside meniscal repair devices, the Mitek RapidLoc (12.5 mumudegrees) and the Smith & Nephew FasT-Fix (curved). The lumen of the femoral artery was injected with barium. Anteroposterior (AP) and lateral radiographs were obtained to ensure visualization of the popliteal artery and its anatomic location. Both devices were inserted into the posterior horn of the lateral meniscus up to the hub of the needle under direct arthroscopic visualization. The distance between the needle and the artery was measured on the radiographs using calipers. The average distance from the needle to the popliteal artery was 0.5 mm (0.5 +/- 0.8 mm, range: 0-2 mm) on the lateral radiograph and 6.0 mm (6.0 +/- 6.2 mm, range: 0-19 mm) on the AP radiograph using the FasT-Fix system without the penetration limiter. None of the RapidLoc needles were within 20 mm of the popliteal artery on either of the radiographs. The FasT-Fix device came within 3 mm of the artery on both AP and lateral radiographs in 43% of the specimens. The risk to the popliteal artery was significantly greater using the FasT-Fix device compared to the RapidLoc device (P < .05). Although both instruments appear safe when used properly, the RapidLoc device seems to be safer with less risk to the popliteal artery than the current FasT-Fix meniscal repair device. When using the FasT-Fix device, the penetration limiter must be used to avoid the complications found in this study.

Entities:  

Mesh:

Year:  2007        PMID: 17665787     DOI: 10.1055/s-0030-1248049

Source DB:  PubMed          Journal:  J Knee Surg        ISSN: 1538-8506            Impact factor:   2.757


  11 in total

Review 1.  Complications in brief: meniscus repair.

Authors:  F Winston Gwathmey; S Raymond Golish; David R Diduch
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

2.  The circumferential compression stitch for meniscus repair.

Authors:  Justin D Saliman
Journal:  Arthrosc Tech       Date:  2013-07-12

3.  Vascular safety during arthroscopic all-inside meniscus suture.

Authors:  Akinobu Nishimura; Aki Fukuda; Ko Kato; Kohzo Fujisawa; Atsumasa Uchida; Akihiro Sudo
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-11-13       Impact factor: 4.342

4.  Inside-Out Repair for Radial Meniscus Tears.

Authors:  Ayoosh Pareek; Michael P O'Malley; Bruce A Levy; Michael J Stuart; Aaron J Krych
Journal:  Arthrosc Tech       Date:  2016-07-25

5.  Clinical results of arthroscopic all-inside lateral meniscal repair using the Meniscal Viper Repair System.

Authors:  Tetsuo Hagino; Satoshi Ochiai; Yoshiyuki Watanabe; Shinya Senga; Masanori Wako; Takashi Ando; Eiichi Sato; Hirotaka Haro
Journal:  Eur J Orthop Surg Traumatol       Date:  2012-11-23

6.  Inside-out meniscus repair.

Authors:  Clay G Nelson; Kevin F Bonner
Journal:  Arthrosc Tech       Date:  2013-11-01

Review 7.  Meniscus tear surgery and meniscus replacement.

Authors:  Javier Vaquero; Francisco Forriol
Journal:  Muscles Ligaments Tendons J       Date:  2016-05-19

8.  Determination of the safe penetration depth during all-inside meniscal repair of the posterior part of the lateral meniscus using the FasT-Fix suture repair system.

Authors:  Mohamed Mahmoud Abouheif; Hayatoshi Shibuya; Takuya Niimoto; Wirat Kongcharoensombat; Masataka Deie; Nobuo Adachi; Mitsuo Ochi
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-04-06       Impact factor: 4.342

9.  Intralaryngeal thyroarytaenoid lateralisation using the Fast-Fix 360 system: a canine cadaveric study.

Authors:  Ludo Stegen; Adriaan M Kitshoff; Bart Van Goethem; Peter Vandekerckhove; Hilde de Rooster
Journal:  Vet Rec Open       Date:  2015-07-27

10.  An arthroscopic repair technique for meniscal tear using a needle and suture: outside-in transfer all-inside repair.

Authors:  Zhiqiang Wang; Yan Xiong; Xin Tang; Qi Li; Zhong Zhang; Jian Li; Gang Chen
Journal:  BMC Musculoskelet Disord       Date:  2019-12-23       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.