H Liang1, Y Yang, K Yang, Y Wu, J M Boone, S R Cherry.
Abstract
A microCT scanner was designed, fabricated and integrated with a previously reported microPET II scanner (Tai et al 2003 Phys. Med. Biol. 48 1519, Yang et al 2004 Phys. Med. Biol. 49 2527), forming a dual modality system for in vivo anatomic and molecular imaging of the mouse. The system was designed to achieve high-spatial-resolution and high-sensitivity PET images with adequate CT image quality for anatomic localization and attenuation correction with low x-ray dose. The system also has relatively high throughput for screening, and a flexible gantry and user interface. X-rays were produced by a 50 kVp, 1.5 mA fixed tungsten anode tube, with a focal spot size of 70 microm. The detector was a 5 x 5 cm(2) photodiode detector incorporating 48 microm pixels on a CMOS array and a fast gadolinium oxysulfide (GOS) intensifying screen. The microCT system has a flexible C-arm gantry design with adjustable detector positioning, which acquires CT projection images around the common microPET/CT bed. The design and the initial characterization of the microCT system is described, and images of the first mouse scans with microPET/CT scanning protocols are shown.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17664583 DOI: 10.1088/0031-9155/52/13/015
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609