Evaluation of Verisyse and Artiflex phakic intraocular lenses during accommodation using Visante optical coherence tomography.

José Luis Güell1, Merce Morral, Oscar Gris, Javier Gaytan, Maite Sisquella, Felicidad Manero.   

Abstract

PURPOSE: To perform a dynamic study of the relationship between Verisyse (AMO) and Artiflex (Ophtec B.V.) phakic intraocular lenses (pIOLs) and anterior chamber structures during accommodation using optical coherence tomography (OCT) (Visante, Carl Zeiss Meditec, Inc.)
SETTING: Institutional practice.
METHODS: Eleven myopic patients were randomly selected to have implantation of a Verisyse pIOL in 1 eye and an Artiflex pIOL in the other. Using a 2-dimensional image, dynamic measurements of the relationship between the anterior surface of the pIOL and the corneal endothelium, the posterior surface of the pIOL and the anterior surface of the crystalline lens, and the pupil diameter were performed using Visante OCT. Physiological accommodation was stimulated by adding lenses in 1.00 diopter (D) steps from +1.00 to -7.00 D.
RESULTS: Both groups had a significant decrease in pupil diameter (P<.0001, generalized linear model [GLM]) and in the distance between the anterior surface of the pIOL and the corneal endothelium (P<.0001, GLM) with accommodation. There were no statistically significant changes in the distance between the posterior surface of either pIOL and the anterior surface of the crystalline lens (P = .2845, GLM). There were no statistically significant differences between the 2 pIOLs in any measurement (P>.05, GLM).
CONCLUSIONS: The results fit with Helmholtz' theory of accommodation as forward movement of the diaphragm iris-crystalline lens was seen. There was a decrease in the distance between the pIOL and corneal endothelium and in the pupil diameter, whereas the distance between both pIOLs and the crystalline lens remained constant throughout the accommodation examination. This suggests that the risk for cataract from intermittent contact between the crystalline lens and IOL from accommodative effort is unlikely.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17662431     DOI: 10.1016/j.jcrs.2007.04.026

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  7 in total

1.  Position-dependent accommodative shift of retropupillary fixated iris-claw lenses.

Authors:  K Schöpfer; A Berger; C Korb; B M Stoffelns; N Pfeiffer; W Sekundo
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-04-19       Impact factor: 3.117

2.  Full anterior segment biometry with extended imaging range spectral domain optical coherence tomography at 1340 nm.

Authors:  Peng Li; Murray Johnstone; Ruikang K Wang
Journal:  J Biomed Opt       Date:  2014-04       Impact factor: 3.170

3.  Lens thickness with age and accommodation by optical coherence tomography.

Authors:  Kathryn Richdale; Mark A Bullimore; Karla Zadnik
Journal:  Ophthalmic Physiol Opt       Date:  2008-09       Impact factor: 3.117

4.  Optical Coherence Tomography Reveals Sigmoidal Crystalline Lens Changes during Accommodation.

Authors:  George A Gibson; Fiona E Cruickshank; James S Wolffsohn; Leon N Davies
Journal:  Vision (Basel)       Date:  2018-08-21

5.  Stability of cachet phakic intraocular lens position during 6-months follow-up.

Authors:  D Schiano Lomoriello; M Lombardo; L Gualdi; L Iacobelli; L Tranchina; P Ducoli; S Serrao
Journal:  Open Ophthalmol J       Date:  2013-03-22

6.  Endothelial distance after phakic iris-fixated intraocular lens implantation: a new safety reference.

Authors:  Tiago Bravo Ferreira; Joana Portelinha
Journal:  Clin Ophthalmol       Date:  2014-01-17

7.  Using a slit lamp-mounted digital high-speed camera for dynamic observation of phakic lenses during eye movements: a pilot study.

Authors:  Martin Alexander Leitritz; Focke Ziemssen; Karl Ulrich Bartz-Schmidt; Bogomil Voykov
Journal:  Clin Ophthalmol       Date:  2014-07-18
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.