Literature DB >> 17661239

[Expiratory pressure reduction (C-Flex Method) versus fix CPAP in the therapy for obstructive sleep apnoea].

M Wenzel1, J Kerl, D Dellweg, T Barchfeld, G Wenzel, D Köhler.   

Abstract

INTRODUCTION: The REM star C-Flex (Fa. Respironics) was introduced in 2003. In contrast to the conventionel fix CPAP mode, the C-Flex mode is characterised by a pressure reduction at the beginning of expiration. In a randomised cross-over design, we investigated if this C-Flex-mode has advantages compared to the fix CPAP mode in terms of treatment quality and patient satisfaction.
METHODS: In this prospective randomised single-blinded cross-over study we investigated 20 patients with obstructive sleep apnoea treated with fix CPAP versus treatment with the C-Flex mode (level 2) for 6 weeks, respectively. We compared the polysomnographically measured quality of treatment and the subjective satisfaction of either form of therapy. Additionally, we measured long-term compliance of the C-Flex therapy in a 3-year follow-up with the integrated counter of the C-Flex devices.
RESULTS: The mean levels of pressure of the fix CPAP and the C-Flex therapies were 8.4 +/- 2.9 mbar in both groups. There was an identical quality of treatment in terms of respiratory events, arousal index, slow wave sleep and Epworth sleepiness scale. The compliance of nocturnal use of the C-Flex and the fix CPAP was identical (6.0 +/- 0.67 C-Flex use vs. 5.8 +/- 0.98 CPAP use [h/night]). The subjective satisfaction was higher in the C-Flex mode at the end of the study since 18 of 20 patients (90%) subjectively prefered the C-Flex mode because of the easier expiration. 19 patients received a C-Flex device for long-term therapy. The 3-year-follow-up showed a regular utilisation of the C-Flex by 16 of 19 (84.2%) of these patients (mean nocturnal use 6.0 +/- 0.9 h/night). 3 of the 19 patients (15.8%) did not use their C-Flex regulary. None of the patients has terminated therapy completely.
CONCLUSION: C-Flex mode and the conventional fix CPAP therapies show an equivalent treatment quality according to polysomnographic data. The expiratory pressure reduction compared to conventional CPAP was felt to be more comfortable by 90% of patients. The long-term-compliance as measured by regular use of the C-Flex device was 84.2% (16 out of 19 patients) after 3 years, these numbers are higher than published data on compliance with conventional CPAP therapy.

Entities:  

Mesh:

Year:  2007        PMID: 17661239     DOI: 10.1055/s-2007-980075

Source DB:  PubMed          Journal:  Pneumologie        ISSN: 0934-8387


  7 in total

1.  Randomized controlled trial comparing flexible and continuous positive airway pressure delivery: effects on compliance, objective and subjective sleepiness and vigilance.

Authors:  Jessie Bakker; Angela Campbell; Alister Neill
Journal:  Sleep       Date:  2010-04       Impact factor: 5.849

Review 2.  Canadian Thoracic Society 2011 guideline update: diagnosis and treatment of sleep disordered breathing.

Authors:  John Fleetham; Najib Ayas; Douglas Bradley; Michael Fitzpatrick; Thomas K Oliver; Debra Morrison; Frank Ryan; Frederick Series; Robert Skomro; Willis Tsai
Journal:  Can Respir J       Date:  2011 Jan-Feb       Impact factor: 2.409

3.  The supraglottic effect of a reduction in expiratory mask pressure during continuous positive airway pressure.

Authors:  Maria J Masdeu; Amit V Patel; Vijay Seelall; David M Rapoport; Indu Ayappa
Journal:  Sleep       Date:  2012-02-01       Impact factor: 5.849

4.  Positive airway pressure initiation: a randomized controlled trial to assess the impact of therapy mode and titration process on efficacy, adherence, and outcomes.

Authors:  Clete A Kushida; Richard B Berry; Alexander Blau; Tami Crabtree; Ingo Fietze; Meir H Kryger; Samuel T Kuna; G Vernon Pegram; Thomas Penzel
Journal:  Sleep       Date:  2011-08-01       Impact factor: 5.849

5.  Longitudinal comparison study of pressure relief (C-Flex) vs. CPAP in OSA patients.

Authors:  Diana C Dolan; Renata Okonkwo; Florian Gfullner; J Randall Hansbrough; Richard J Strobel; Leon Rosenthal
Journal:  Sleep Breath       Date:  2008-06-13       Impact factor: 2.816

6.  Assessment of the impact on compliance of a new CPAP system in obstructive sleep apnea.

Authors:  Alison J Wimms; Glenn N Richards; Adam V Benjafield
Journal:  Sleep Breath       Date:  2013-03       Impact factor: 2.816

7.  Pressure modification or humidification for improving usage of continuous positive airway pressure machines in adults with obstructive sleep apnoea.

Authors:  Barry Kennedy; Toby J Lasserson; Dariusz R Wozniak; Ian Smith
Journal:  Cochrane Database Syst Rev       Date:  2019-12-02