Accurate flip-angle calibration for 13C MRI.

Ileana Hancu1, Ronald Watkins1, Susan J Kohler2, Richard P Mallozzi1.   

Abstract

(13)C imaging and spectroscopy in the presence of injected labeled compounds can vastly extend the capability of MRI to perform metabolic imaging. The details of imaging in the presence of injected compounds, however, pose new technological challenges. Pulse sequences, in general, rely on precise flip-angle (FA) calibration to create high signal-to-noise ratio (SNR), artifact-free images. Signal quantification also requires precise knowledge of the excitation FA. In MRI scans that rely on signal acquisitions from injected compounds, however, such FA calibration is challenged by low natural-abundance (13)C signal levels before injection, and by time-varying signal following injection. A method to precisely set the FA at the (13)C frequency based on FA calibration at the (23)Na frequency is presented here. A practical implementation of a coil (a dual-tuned, (23)Na/(13)C low-pass birdcage coil) suitable for this calibration in vivo is also documented. Accurate FA calibration is demonstrated at the (13)C frequency for in vivo rat experiments using this approach. (c) 2007 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17659619     DOI: 10.1002/mrm.21252

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   3.737


  2 in total

1.  Measuring B1 distributions by B1 phase encoding.

Authors:  Kalina V Jordanova; Dwight G Nishimura; Adam B Kerr
Journal:  Magn Reson Med       Date:  2016-01-17       Impact factor: 4.668

Review 2.  Acquisition strategies for spatially resolved magnetic resonance detection of hyperpolarized nuclei.

Authors:  Geoffrey J Topping; Christian Hundshammer; Luca Nagel; Martin Grashei; Maximilian Aigner; Jason G Skinner; Rolf F Schulte; Franz Schilling
Journal:  MAGMA       Date:  2019-12-06       Impact factor: 2.310

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.