| Literature DB >> 17645996 |
Jennifer S Huber, William W Moses, Gin-Chung Wang, Stephen E Derenzo, Ronald H Huesman, Jinyi Qi, Patrick Virador, Woon-Seng Choong, Emanuele Mandelli, Eric Beuville, Marzio Pedrali-Noy, Brad Krieger, Gerrit J Meddeler.
Abstract
We present a retrospective on the LBNL Positron Emission Mammography (PEM) project, looking back on our design and experiences. The LBNL PEM camera utilizes detector modules that are capable of measuring depth of interaction (DOI) and places them into 4 detector banks in a rectangular geometry. In order to build this camera, we had to develop the DOI detector module, LSO etching, Lumirror-epoxy reflector for the LSO array (to achieve optimal DOI), photodiode array, custom IC, rigid-flex readout board, packaging, DOI calibration and reconstruction algorithms for the rectangular camera geometry. We will discuss the high-lights (good and bad) of these developments.Year: 2006 PMID: 17645996 DOI: 10.1016/S1120-1797(06)80026-8
Source DB: PubMed Journal: Phys Med ISSN: 1120-1797 Impact factor: 2.685