3D C-arm as an alternative modality to CT in postmortem imaging: technical feasibility.

Philipp Pohlenz1, Marco Blessmann, Lars Oesterhelweg, Christan R Habermann, Philipp G C Begemann, Christian Schmidgunst, Felix Blake, Dirk Schulze, Klaus Püschel, Rainer Schmelzle, Max Heiland.   

Abstract

OBJECT: The aim of our study was to demonstrate the image quality of the new device using human cadavers, extending the horizon of available imaging modalities in forensic medicine.
MATERIALS AND METHODS: Six human cadavers were examined, revealing C-arm data sets of the head, neck thorax, abdomen and pelvis. High-resolution mode was performed with 500 fluoroscopy shots during a 190 degrees orbital movement with a constant tube voltage of 100 kV and a current of 4.6 mA. Based on these data sets subsequent three-dimensional reconstructions were generated.
RESULTS: Reconstructed data sets revealed high-resolution images of all skeletal structures in a near-CT quality. The same image quality was available in all reconstruction planes. Artefacts caused by restorative dental materials are less accentuated in CBCT data sets. The system configuration was not powerful enough to generate sufficient images of intracranial structures.
CONCLUSION: After the here-demonstrated encouraging preliminary results, the forensic indications that would be suitable for imaging with a 3D C-arm have to be defined. Promising seems the visualization local limited region of interest as the cervical spine or the facial skeleton.

Entities:  

Mesh:

Year:  2007        PMID: 17640838     DOI: 10.1016/j.forsciint.2007.06.005

Source DB:  PubMed          Journal:  Forensic Sci Int        ISSN: 0379-0738            Impact factor:   2.395


  2 in total

1.  [C-arm computed tomography for transarterial chemoperfusion and chemo-embolization of thoracic lesions].

Authors:  T J Vogl; N N Naguib; N-E Nour-Eldin; T Lehnert; E Mbalisike
Journal:  Radiologe       Date:  2009-09       Impact factor: 0.635

Review 2.  Virtopsy: Touch-free autopsy.

Authors:  Raj Kumar Badam; Triekan Sownetha; D B Gandhi Babu; Shefali Waghray; Lavanya Reddy; Komali Garlapati; Sunanda Chavva
Journal:  J Forensic Dent Sci       Date:  2017 Jan-Apr
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.