| Literature DB >> 17629916 |
Stanley E Rittgers1, Matt T Oberdier, Sharath Pottala.
Abstract
Due to the relatively limited amount of work done to date on developing prosthetic vein (as opposed to cardiac) valves, advances in this topic require progress in three distinct areas: 1) improved device design, 2) relevant device testing conditions, and, 3) appropriate parameters for evaluation of results. It is the purpose of this paper to address two of these issues (#2 and #3) by: 1) performing a study of normal volunteers to quantify the anatomy and hemodynamic features of healthy venous valves, 2) construction of a 2-step, in vitro testing procedure, which simulates both physiologic and postural conditions seen in the lower extremity venous system, and, 3) defining several modified and new parameters which quantify dynamic valve characteristics.Entities:
Mesh:
Year: 2007 PMID: 17629916 PMCID: PMC1947988 DOI: 10.1186/1475-925X-6-29
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Figure 1A schematic of the vertical column 2:1 Mock-Up System.
Figure 2A schematic of the horizontal 1:1 Pulsatile Flow System.
1:1 Pulsatile Flow System Testing Protocol.
| 0.35 | 15 | OFF | |
| 0.75 | 30 | OFF | |
| 0.35 | 15 | ON | |
| 0.75 | 30 | ON |
Figure 3Prototype design of venous valve in the Open (left) and Closed (right) positions consisting of a solid circular base, frame struts with uniquely shaped flanges, and a flexible leaflet material.
Figure 4Actual 2:1 scale device with optimized frame and shaped Biospan® leaflet. A threaded extension from the base (not shown) screwed into the test section fixture to improve alignment and prevent leakage.
1:1 Pulsatile Flow Results for Prototype vs. 'No Device' in Supine Position.
| None | 13.3 ± 0.0272 | 21.3 ± 0.271 | 1.23* | |
| Prototype | 14.6 ± 0.0012 | 23.5 ± 0.0338 | 1.12 ± 0.0166 | |
| Change | 1.33 | 2.23 | 0.11 | |
| None | 7.97 ± 0.0935 | 6.00 ± 0.163 | 1.23* | |
| Prototype | 8.81 ± 0.157 | 7.37 ± 0.110 | 1.09 ± 0.0111 | |
| Change | 1.84 | 1.37 | 0.14 |
* Calculated value based on tube dimensions
1:1 Pulsatile Flow Results for Prototype vs. 'No Device' in Standing Position.
| None | 59.6 ± 0.830 | 205 ± 5.37 | 283 ± 7.93 | -319 ± 22.1 | |
| Prototype | 5.83 ± 0.688 | 18.1 ± 2.12 | 2.73 ± 0.654 | 36.7 ± 0.785 | |
| None | 7.35 ± 0.28 | 24.6 ± 0.859 | 4.30 ± 0.333 | 93.9 ± 0.429 | |
| Prototype | 2.21 ± 0.978 | 7.10 ± 3.27 | 0.451 ± 0.411 | 99.4 ± 0.562 |
1:1 Pulsatile Flow Results for Repeat Measurements in Supine Position.
| 1st | 1.33 ± 0.0012 | 2.23 ± 0.0338 | 1.12 ± 0.0166 | |
| 2nd | 1.41 ± 0.0221* | 2.29 ± 0.0379* | 1.16 ± 0.015* | |
| Change | 0.08 | 0.06 | 0.04 | |
| 1st | 1.84 ± 0.151 | 1.37 ± 0.110 | 1.09 ± 0.0111 | |
| 2nd | 2.33 ± 0.0926* | 1.73 ± 0.0889* | 1.08 ± 0.0215 | |
| Change | 0.49 | 0.36 | 0.01 |
1 Relative to 'No Device' value
* p < 0.05 vs. 1st Test Run
1:1 Pulsatile Flow Results for Repeat Measurements in Standing Position.
| 1st | 5.83 ± 0.688 | 18.1± 2.12 | 2.73 ± 0.654 | 36.7 ± 0.785 | |
| 2nd | 3.99 ± 0.537* | 12.7 ± 2.07* | 1.29 ± 0.371* | 38.4 ± 0.52* | |
| Change | 1.84 | 5.4 | 1.44 | 1.7 | |
| 1st | 2.21 ± 0.978 | 7.10 ± 3.27 | 0.451 ± 0.411 | 99.4 ± 0.562 | |
| 2nd | 1.55 ± 1.74 | 4.75 ± 5.27 | 0.384 ± 0.375 | 99.6 ± 0.443 | |
| Change | 0.66 | 2.35 | 0.067 | 0.2 |
* p < 0.05 vs. 1st Test Run