Detection of carbon monoxide production as a result of the interaction of five volatile anesthetics and desiccated sodalime with an electrochemical carbon monoxide sensor in an anesthetic circuit compared to gas chromatography.

Christiaan Keijzer1, Roberto S G M Perez, Jaap J de Lange.   

Abstract

OBJECTIVES: There is a continuing risk of production of toxic levels of carbon monoxide (CO) as a result of interaction of volatile anesthetics and desiccated strong base carbon dioxide absorbents like soda lime. The aim of this study is to establish the reliability of detection of CO levels by an electrochemical carbon monoxide sensor compared to gas chromatography.
METHODS: Completely desiccated sodalime was conducted through a circle anesthesia system connected to an artificial lung. For different rates of CO production, a low flow anesthesia with a oxygen/nitrous oxide mixture was maintained using five volatile anesthetics. For quantification of CO production, a portable gas chromatograph (GC) was connected to this setup, as well as a Bedfont EC40 electrochemical carbon monoxide sensor (ES) with a claimed reliable sensitivity of 0-200 parts per million (ppm) and a maximum detection range of more than 5500 ppm. To assess the agreement between the GC and ES measurements the intra class correlation coefficient (ICC) and the 95% limits of agreement were calculated. Bland and Altman scatterplots were made to visualize the difference between measurements.
RESULTS: For concentrations up to 200 ppm, no significant differences between the GC and ES mean CO measurements were found in the halothane experiments. However CO was not accurately measured at every moment during these experiments by the ES. For concentrations above 200 ppm the results of the two instruments differed significantly. The ES malfunctioned when exposed to sevoflurane and desiccated sodalime.
CONCLUSIONS: From these data we conclude that the ES can only be used as an indicator of CO production. When this sensor is used with sevoflurane and desiccated sodalime it is not capable of normal operation. The use of a strong base free carbon dioxide absorbent is therefore recommended.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17597416     DOI: 10.1007/s10877-007-9083-y

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  12 in total

1.  Investigation of electrochemical carbon monoxide sensor monitoring of anesthetic gas mixtures.

Authors:  Joseph A Bermudez
Journal:  Anesthesiology       Date:  2003-11       Impact factor: 7.892

2.  Production of compound A and carbon monoxide in circle systems: an in vitro comparison of two carbon dioxide absorbents.

Authors:  M M R F Struys; M P L A Bouche; G Rolly; Y D I Vandevivere; D Dyzers; W Goeteyn; L F M Versichelen; J F P Van Bocxlaer; E P Mortier
Journal:  Anaesthesia       Date:  2004-06       Impact factor: 6.955

3.  Acute respiratory distress syndrome after an exothermic Baralyme-sevoflurane reaction.

Authors:  Randy S Fatheree; Barbara L Leighton
Journal:  Anesthesiology       Date:  2004-08       Impact factor: 7.892

4.  Absorbents differ enormously in their capacity to produce compound A and carbon monoxide.

Authors:  C R Stabernack; R Brown; M J Laster; R Dudziak; E I Eger
Journal:  Anesth Analg       Date:  2000-06       Impact factor: 5.108

5.  Severe carbon monoxide poisoning during desflurane anesthesia.

Authors:  P D Berry; D I Sessler; M D Larson
Journal:  Anesthesiology       Date:  1999-02       Impact factor: 7.892

Review 6.  AANA journal course: update for nurse anesthetists--carbon monoxide poisoning: role of the anesthesia machine's carbon dioxide absorption system.

Authors:  C Altman
Journal:  AANA J       Date:  1996-02

7.  Dry soda lime markedly degrades sevoflurane during simulated inhalation induction.

Authors:  W Funk; M Gruber; K Wild; J Hobbhahn
Journal:  Br J Anaesth       Date:  1999-02       Impact factor: 9.166

8.  Amsorb: a new carbon dioxide absorbent for use in anesthetic breathing systems.

Authors:  J M Murray; C W Renfrew; A Bedi; C B McCrystal; D S Jones; J P Fee
Journal:  Anesthesiology       Date:  1999-11       Impact factor: 7.892

9.  Performance of an electrochemical carbon monoxide monitor in the presence of anesthetic gases.

Authors:  M Dunning; H J Woehlck
Journal:  J Clin Monit       Date:  1997-11

10.  The elimination of sodium and potassium hydroxides from desiccated soda lime diminishes degradation of desflurane to carbon monoxide and sevoflurane to compound A but does not compromise carbon dioxide absorption.

Authors:  M A Neumann; M J Laster; R B Weiskopf; D H Gong; R Dudziak; H Förster; E I Eger
Journal:  Anesth Analg       Date:  1999-09       Impact factor: 5.108

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.