Literature DB >> 15144299

Production of compound A and carbon monoxide in circle systems: an in vitro comparison of two carbon dioxide absorbents.

M M R F Struys1, M P L A Bouche, G Rolly, Y D I Vandevivere, D Dyzers, W Goeteyn, L F M Versichelen, J F P Van Bocxlaer, E P Mortier.   

Abstract

Two new generation carbon dioxide absorbents, DrägerSorb Free and Amsorb Plus, were studied in vitro for formation of compound A or carbon monoxide, during minimal gas flow (500 ml x min(-1)) with sevoflurane or desflurane. Compound A was assessed by gas chromatography/mass spectrometry and carbon monoxide with continuous infrared spectrometry. Fresh and dehydrated absorbents were studied. Mean (SD) time till exhaustion (inspiratory carbon dioxide concentration >or= 1 kPa) with fresh absorbents was longer with DrägerSorb Free (1233 (55) min) than with Amsorb Plus (1025 (55) min; p < 0.01). For both absorbents, values of compound A were < 1 ppm and therefore below clinically significant levels, but were up to 0.25 ppm higher with DrägerSorb Free than with Amsorb Plus. Using dehydrated absorbents, values of compound A were about 50% lower than with fresh absorbents and were identical for DrägerSorb Free and Amsorb Plus. With dehydrated absorbents, no detectable carbon monoxide was found with desflurane.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15144299     DOI: 10.1111/j.1365-2044.2004.03704.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  2 in total

1.  Important role of calcium chloride in preventing carbon monoxide generation during desflurane degradation with alkali hydroxide-free carbon dioxide absorbents.

Authors:  Takahiro Ando; Atsushi Mori; Rie Ito; Kimitoshi Nishiwaki
Journal:  J Anesth       Date:  2017-08-22       Impact factor: 2.078

2.  Detection of carbon monoxide production as a result of the interaction of five volatile anesthetics and desiccated sodalime with an electrochemical carbon monoxide sensor in an anesthetic circuit compared to gas chromatography.

Authors:  Christiaan Keijzer; Roberto S G M Perez; Jaap J de Lange
Journal:  J Clin Monit Comput       Date:  2007-06-28       Impact factor: 2.502

  2 in total