| Literature DB >> 17587662 |
J P Hill1, D S Coburn, Y J Kim, T Gog, D M Casa, C N Kodituwakku, H Sinn.
Abstract
The design and commissioning of an inelastic X-ray scattering instrument at CMC-XOR at the Advanced Photon Source is reported. The instrument features a 2 m vertical-scattering arm with a novel counterweight design to reduce the twisting moment as the arm is moved in the scattering plane. A Ge(733) spherical analyzer was fabricated and an overall resolution of 118 meV (FWHM) was obtained with a Si(444) monochromator and a Si(111) pre-monochromator. Early results from a representative cuprate, La(2)CuO(4), are reported.Entities:
Year: 2007 PMID: 17587662 DOI: 10.1107/S0909049507018006
Source DB: PubMed Journal: J Synchrotron Radiat ISSN: 0909-0495 Impact factor: 2.616