The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration.

Berthold Bein1, Patrick Meybohm, Erol Cavus, Jochen Renner, Peter H Tonner, Markus Steinfath, Jens Scholz, Volker Doerges.   

Abstract

BACKGROUND: Reliable measurement of cardiac output (CO) is important in the critically ill. Pulse contour-derived CO (PCCO) has been evaluated during stable hemodynamics, but is sensitive to changes in vascular tone and has not been validated under conditions of changing hemodynamics. Furthermore, PCCO requires calibration for the individual vascular impedance by transpulmonary thermodilution CO (TPCO), and the required frequency of recalibration to maintain accurate measurements, especially during changing conditions, has not been confirmed. We compared PCCO measurements of CO with TPCO and continuous and bolus pulmonary artery CO (CCO and BCO, respectively) during conditions of uncontrolled hemorrhage and resuscitation with norepinephrine.
METHODS: Thirteen pigs were anesthetized and instrumented for determination of CO by BCO and CCO, respectively, as well as bolus TPCO and PCCO. Uncontrolled hemorrhage was accomplished by liver incision. When mean arterial blood pressure was <25 mm Hg, or heart rate declined progressively to <20% of its peak value, vasopressor therapy was started. TPCO and BCO were performed after induction of anesthesia and 15 min after start of therapy, and PCCO and CCO were obtained repeatedly. CO measurements were compared using Bland-Altman analysis.
RESULTS: Mean arterial blood pressure, CO and systemic vascular resistance decreased after hemorrhage (P < 0.001 and <0.01, respectively). Bias and limits of agreement between CCO and PCCO (0.54 L/min; 1.46 L/min) increased after hemorrhage (-3.49; 6.12) and further deteriorated after norepinephrine administration (-8.01; 9.9). After recalibration, bias and limits of agreement returned to -0.51 and 1.28.
CONCLUSIONS: PCCO needs frequent recalibration during hemorrhage and after vasopressor administration.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17578965     DOI: 10.1213/01.ane.0000268140.02147.ed

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  20 in total

Review 1.  Continuous and less invasive central hemodynamic monitoring by blood pressure waveform analysis.

Authors:  Ramakrishna Mukkamala; Da Xu
Journal:  Am J Physiol Heart Circ Physiol       Date:  2010-07-09       Impact factor: 4.733

Review 2.  Advanced hemodynamic monitoring: principles and practice in neurocritical care.

Authors:  Christos Lazaridis
Journal:  Neurocrit Care       Date:  2012-02       Impact factor: 3.210

3.  Correlation of thermodilution-derived extravascular lung water and ventilation/perfusion-compartments in a porcine model.

Authors:  Erik K Hartmann; Bastian Duenges; James E Baumgardner; Klaus Markstaller; Matthias David
Journal:  Intensive Care Med       Date:  2013-04-18       Impact factor: 17.440

4.  Perioperative management of the pediatric cardiac transplantation patient.

Authors:  Avihu Z Gazit; James Fehr
Journal:  Curr Treat Options Cardiovasc Med       Date:  2011-10

Review 5.  Applied physiology at the bedside to drive resuscitation algorithms.

Authors:  Andre L Holder; Michael R Pinsky
Journal:  J Cardiothorac Vasc Anesth       Date:  2014-12       Impact factor: 2.628

6.  Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices.

Authors:  Xavier Monnet; Nadia Anguel; Brice Naudin; Julien Jabot; Christian Richard; Jean-Louis Teboul
Journal:  Crit Care       Date:  2010-06-10       Impact factor: 9.097

7.  Are management decisions in critical patients changed with use of hemodynamic parameters from transpulmonary thermodilution technique?

Authors:  Ye Rim Chang; Seok Ho Choi; Sung Wook Chang
Journal:  Ann Transl Med       Date:  2019-08

8.  Effect of norepinephrine dosage and calibration frequency on accuracy of pulse contour-derived cardiac output.

Authors:  Matthias Gruenewald; Patrick Meybohm; Jochen Renner; Ole Broch; Amke Caliebe; Norbert Weiler; Markus Steinfath; Jens Scholz; Berthold Bein
Journal:  Crit Care       Date:  2011-01-17       Impact factor: 9.097

9.  Variations in arterial blood pressure are associated with parallel changes in FlowTrac/Vigileo-derived cardiac output measurements: a prospective comparison study.

Authors:  Savvas Eleftheriadis; Zisis Galatoudis; Vasilios Didilis; Ioannis Bougioukas; Julika Schön; Hermann Heinze; Klaus-Ulrich Berger; Matthias Heringlake
Journal:  Crit Care       Date:  2009-11-09       Impact factor: 9.097

10.  A comparison of third-generation semi-invasive arterial waveform analysis with thermodilution in patients undergoing coronary surgery.

Authors:  Ole Broch; Jochen Renner; Matthias Gruenewald; Patrick Meybohm; Jan Schöttler; Markus Steinfath; Manu Malbrain; Berthold Bein
Journal:  ScientificWorldJournal       Date:  2012-07-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.