Literature DB >> 17575578

Fully automated analysis of Carbohydrate-Deficient Transferrin (CDT) by using a multicapillary electrophoresis system.

Federica Bortolotti1, Giorgia De Paoli, Jennifer P Pascali, Franco Tagliaro.   

Abstract

BACKGROUND: The techniques universally believed as most reliable for determination of Carbohydrate-Deficient Transferrin (CDT) are HPLC and capillary electrophoresis (CE). Recently a reagent kit for CDT analysis to be used in a multicapillary electropherograph has been introduced. The present work was aimed at a validation of this new commercial system by application of the analytical chemistry parameters and by comparison with validated CE and HPLC methods.
METHODS: One hundred forty serum samples were analyzed with multicapillary CE (Capillarys, Sebia, France) using kit reagents (CAPILLARYS CDT assay), with single capillary electropherograph (P/ACE MDQ, Beckman Coulter, USA) using an original method developed by our group and with HPLC, performed on a gradient HPLC (Shimadzu Europe, Germany), using column and reagents provided in kit (ClinRep, Recipe, Germany).
RESULTS: The separation efficiency of the multicapillary system was about 15,000 theoretical plates/column. The resolution of the transferrin isoforms ranged from 1.23 to 1.67. The variation coefficient was < 10% in both intra-run and between run tests. The correlation of results of multicapillary CE with those obtained with single CE and HPLC was statistically significant.
CONCLUSIONS: The multicapillary system shows high productivity with good analytical performances; however, a confirmation of positive results with an alternative method is required.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17575578     DOI: 10.1016/j.cca.2006.12.028

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  2 in total

1.  Carbohydrate-Deficient Transferrin Determination in a Clinical Setting: Consistency Between Capillary Electrophoresis Assays and Utility of HPLC as a Confirmatory Test.

Authors:  Agnese Veronesi; Cristina Rota; Tommaso Trenti; Elisabetta Cariani
Journal:  J Clin Lab Anal       Date:  2015-10-26       Impact factor: 2.352

2.  Novel approach to identifying the hepatitis B virus pre-S deletions associated with hepatocellular carcinoma.

Authors:  Zhi-Mei Zhao; Yan Jin; Yu Gan; Yu Zhu; Tao-Yang Chen; Jin-Bing Wang; Yan Sun; Zhi-Gang Cao; Geng-Sun Qian; Hong Tu
Journal:  World J Gastroenterol       Date:  2014-10-07       Impact factor: 5.742

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.