Literature DB >> 17559108

Standardized evaluation and reporting of stent fractures in clinical trials of noncoronary devices.

Michael Jaff1, Michael Dake, Jeffrey Pompa, Gary Ansel, Tony Yoder.   

Abstract

OBJECTIVES: Define the prospective surveillance and reporting parameters for assessment of metallic stents used in clinical trials of superficial femoral artery atherosclerosis.
BACKGROUND: While advances in medical therapies to improve atherosclerotic risk factor profiles in patients with peripheral arterial disease have focused on prevention of disease progression, the development of endovascular devices is altering the therapeutic algorithm for patients with peripheral arterial disease. Stents have emerged as one of the advances that may promote durable patency, particularly in the infrainguinal arteries.
METHODS: The Zilver-PTX is a prospective multicenter randomized trial of bare metal versus drug-eluting nitinol stents for superficial femoral artery peripheral arterial disease. A predefined surveillance program is mandated in this trial. During the genesis of this program, it became clear that no single standard has existed to identify, classify, and report fractures of metallic stents placed in peripheral arteries.
RESULTS: This report defines the key components of prospective stent fracture surveillance, including location of radiographic images obtained, types of radiographic equipment, methods of limb, and image intensifier positioning, intervals required for imaging, grading schemes for reporting, and methods of analyzing the images.
CONCLUSIONS: As the proliferation of minimally invasive revascularization strategies continues, the stability of metallic endoprostheses is increasingly critical. Uniform surveillance of these devices for fracture and disarticulation is important in order to compare the relative merits of various stent-based devices in clinical trials. (c) 2007 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2007        PMID: 17559108     DOI: 10.1002/ccd.21240

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  9 in total

1.  Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section.

Authors:  Kartik V Bulusu; Michael W Plesniak
Journal:  J Vis Exp       Date:  2016-07-19       Impact factor: 1.355

2.  Structural Mechanics Predictions Relating to Clinical Coronary Stent Fracture in a 5 Year Period in FDA MAUDE Database.

Authors:  Kay D Everett; Claire Conway; Gerard J Desany; Brian L Baker; Gilwoo Choi; Charles A Taylor; Elazer R Edelman
Journal:  Ann Biomed Eng       Date:  2015-10-14       Impact factor: 3.934

3.  Stent fractures after superficial femoral artery stenting.

Authors:  Jae Young Park; Yong Sun Jeon; Soon Gu Cho; Chan Ik Jin; Kyung Rae Kim; Woo Young Shin; Jang Yong Kim; Kee Chun Hong
Journal:  J Korean Surg Soc       Date:  2012-08-27

4.  Fracture of zotarolimus-eluting stent after implantation.

Authors:  Meng-Chieh Wu; Chen-Chuan Cheng; Tsuei-Yuan Huang
Journal:  Tex Heart Inst J       Date:  2009

5.  Large left ventricular non-infectious vegetation in patient with eosinophilic granulomatosis with polyangiitis.

Authors:  Yun-Seok Song; Sang-Hoon Seol; Jino Park; Dong-Kie Kim; Yeo-Jeong Song; Seunghwan Kim; Ki-Hun Kim; Doo-Il Kim; Chan-Seon Park; Yeon-Mi Kim
Journal:  Cardiovasc J Afr       Date:  2020-01-29       Impact factor: 1.167

6.  Clinical characteristics of coronary drug-eluting stent fracture: insights from a two-center des registry.

Authors:  Kyung Woo Park; Jin Joo Park; In-Ho Chae; Jae-Bin Seo; Han-Mo Yang; Hae-Young Lee; Hyun-Jae Kang; Young-Seok Cho; Tae-Jin Yeon; Woo-Young Chung; Bon-Kwon Koo; Dong-Ju Choi; Byung-Hee Oh; Young-Bae Park; Hyo-Soo Kim
Journal:  J Korean Med Sci       Date:  2010-12-22       Impact factor: 2.153

7.  Hybrid surgery techniques for the treatment of in-stent restenosis after 5 years of femoral artery self-expanding bare-metal stent implantation: A case report.

Authors:  Jianguo Zhou; Guosong Zha; Guosheng Qian
Journal:  Medicine (Baltimore)       Date:  2022-03-11       Impact factor: 1.817

Review 8.  Coronary stent fracture: a recently appreciated phenomenon with clinical relevance.

Authors:  Madjid Chinikar; Parham Sadeghipour
Journal:  Curr Cardiol Rev       Date:  2014-11

9.  Ex vivo assessment and in vivo validation of non-invasive stent monitoring techniques based on microwave spectrometry.

Authors:  Carolina Gálvez-Montón; Gianluca Arauz-Garofalo; Oriol Rodriguez-Leor; Carolina Soler-Botija; Susana Amorós García de Valdecasas; Flavio David Gerez-Britos; Antoni Bayes-Genis; Juan Manuel O'Callaghan; Ferran Macià; Javier Tejada
Journal:  Sci Rep       Date:  2018-10-04       Impact factor: 4.379

  9 in total