Hybrid PET/CT of the thorax: when is computer registration necessary?

Matthew D Gilman1, Alan J Fischman, Vikram Krishnasetty, Elkan F Halpern, Suzanne L Aquino.   

Abstract

OBJECTIVE: To determine if computer registration improves the image fusion of hybrid positron emission tomography (PET)/computed tomography (CT) data sets in the thorax acquired during different breathing maneuvers.
METHODS: Hybrid PET/CT scans were acquired with varying CT respiratory instructions. The scans of 64 patients with 5 different breathing maneuvers were reviewed, including expiration, suspended breath hold, quiet breathing, small breath in, and regular breath in. The PET/CT data sets were reviewed before registration, after linear registration, and combined linear/nonlinear registration. Each PET/CT data set was graded for the quality of alignment at 5 anatomical locations (diaphragm, aortic arch, heart, thoracic spine, and lung apices) and rated from 1 (very poor) to 5 (excellent). The Kruskal-Wallis test was used to compare alignment between the breathing protocols, and Wilcoxon signed rank test was used to compare the registration techniques.
RESULTS: The quality of anatomical alignment between superimposed (nonregistered) PET and attenuation correction CT (CTAC) acquired during expiration, suspended breath hold, and quiet breathing was excellent, with no significant changes in alignment following registration. The quality of anatomical alignment between superimposed (nonregistered) PET and CTAC acquired during inspiration showed significant misalignment at the heart (P = 0.001, P < 0.0001) and diaphragm (P = 0.0001, P < 0.0001) which did not correct with linear registration. Alignment significantly improved with additional nonlinear registration at the diaphragm (P = 0.008, P = 0.0002), although cardiac misalignment remained.
CONCLUSIONS: With hybrid PET/CT, excellent anatomical alignment between PET and CTAC scans can be achieved with CT acquisition during expiration, suspended breath hold, or quiet breathing, and additional software registration is not necessary. Nonlinear registration significantly improves alignment discrepancies at the diaphragm if inspiratory methods are used, however, cardiac misalignment remains.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17538286     DOI: 10.1097/01.rct.0000237817.18678.9c

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  5 in total

1.  Hybrid cardiac imaging: more than the sum of its parts?

Authors:  Oliver Gaemperli; Philipp A Kaufmann
Journal:  J Nucl Cardiol       Date:  2008 Jan-Feb       Impact factor: 5.952

2.  Impact of rigid and nonrigid registration on the determination of 18F-FDG PET-based tumour volume and standardized uptake value in patients with lung cancer.

Authors:  Aleksandar Grgic; Elena Ballek; Jochen Fleckenstein; Norbert Moca; Stephanie Kremp; Andrea Schaefer; Jan-Martin Kuhnigk; Christian Rübe; Carl-Martin Kirsch; Dirk Hellwig
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-01-22       Impact factor: 9.236

3.  Correlative microscopy methods that maximize specimen fidelity and data completeness, and improve molecular localization capabilities.

Authors:  Elizabeth A Smith; Bertrand P Cinquin; Gerry McDermott; Mark A Le Gros; Dilworth Y Parkinson; Hong Tae Kim; Carolyn A Larabell
Journal:  J Struct Biol       Date:  2013-03-24       Impact factor: 2.867

Review 4.  Multimodality imaging: an update on PET/CT technology.

Authors:  Osama Mawlawi; David W Townsend
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-03       Impact factor: 9.236

Review 5.  An Overview on Image Registration Techniques for Cardiac Diagnosis and Treatment.

Authors:  Azira Khalil; Siew-Cheok Ng; Yih Miin Liew; Khin Wee Lai
Journal:  Cardiol Res Pract       Date:  2018-08-08       Impact factor: 1.866

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.