| Literature DB >> 17533529 |
C P Shortt1, N F Fanning, L Malone, J Thornton, P Brennan, M J Lee.
Abstract
PURPOSE: To assess radiation dose to the thyroid in patients undergoing neurointerventional procedures and to evaluate dose reduction to the thyroid by lead shielding. METHODS AND MATERIALS: A randomized patient study was undertaken to evaluate the dose reduction by thyroid lead shields and assess their practicality in a clinical setting. Sixty-five patients attending for endovascular treatment of arteriovenous malformations (AVMs) and aneurysms were randomized into one of 2 groups a) No Thyroid Shield and b) Thyroid Lead Shield. Two thermoluminescent dosimeters (TLDs) were placed over the thyroid gland (1 on each side) at constant positions on each patient in both groups. A thyroid lead shield (Pb eq. 0.5 mm) was placed around the neck of patients in the thyroid lead shield group after the neurointerventional radiologist had obtained satisfactory working access above the neck. The total dose-area-product (DAP) value, number and type of digital subtraction angiography (DSA) runs and fluoroscopy time were recorded for all patients.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17533529 DOI: 10.1007/s00270-007-9093-7
Source DB: PubMed Journal: Cardiovasc Intervent Radiol ISSN: 0174-1551 Impact factor: 2.740