| Literature DB >> 17532496 |
Yuta Shibamoto1, Asaka Naruse, Hiroshi Fukuma, Shiho Ayakawa, Chikao Sugie, Natsuo Tomita.
Abstract
Influences of iodinated contrast media on dose calculation were studied in 26 patients. Mean increases in monitor units by contrast media administration were less than 1% and considered negligible in planning of whole-brain, whole-neck, mediastinal, and whole-pelvic irradiation. However, mean increases over 2% were seen in planning of upper-abdominal radiotherapy.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17532496 DOI: 10.1016/j.radonc.2007.05.015
Source DB: PubMed Journal: Radiother Oncol ISSN: 0167-8140 Impact factor: 6.280