Literature DB >> 24945375

Helical and Static-port Tomotherapy Using the Newly-developed Dynamic Jaws Technology for Lung Cancer.

Yoshihiko Manabe1, Yuta Shibamoto2, Chikao Sugie2, Akihiro Hayashi2, Taro Murai2, Takeshi Yanagi2.   

Abstract

With the newly developed dynamic jaws technology, radiation dose for the cranio-caudal edges of a target can be lowered in the treatment with tomotherapy. We compared dynamic-jaw- and fixed-jaw-mode plans for lung cancer. In 35 patients, four plans using the 2.5-cm dynamic-, 2.5-cm fixed-, 5.0-cm dynamic-, and 5.0-cm fixed-jaw modes were generated. For 10 patients with upper lobe stage I lung cancer, the helical tomotherapy mode was used. Fifty-six Gy in 8 fractions was prescribed as a minimum coverage dose for 95% of the target (D95%). For 25 patients with locally advanced lung cancer, plans using four static ports (TomoDirect® mode) were made. Sixty Gy in 30 daily fractions for the primary tumor and swollen lymph nodes and 51 Gy in 30 fractions for prophylactic lymph node areas were prescribed as median doses. The mean conformity index of the planning target volume were similar among the four plans. The mean V5 Gy of the lung for 2.5-cm dynamic-, 2.5-cm fixed-, 5.0-cm dynamic-, and 5.0-cm fixed-jaw mode plans were 18.5%, 21.8%, 20.1%, and 29.4%, respectively (p < 0.0001), for patients with stage I lung cancer, and 37.3%, 38.7%, 40.4%, and 44.0%, respectively (p < 0.0001), for patients with locally advanced lung cancer. The mean V5 Gy of the whole body was 1,826, 2,143, 1,983, and 2,939 ml, respectively (p < 0.0001), for patients with stage I lung cancer and 4,849, 5,197, 5,220, and 6,154 ml, respectively (p < 0.0001), for patients with locally advanced lung cancer. Treatment time was reduced by 21-39% in 5.0-cm dynamic-jaw plans compared to 2.5-cm plans. Regarding dose distribution, 2.5-cm dynamic-jaw plans were the best, and 5.0-cm dynamic-jaw plans were comparable to 2.5-cm fixed-jaw plans with shorter treatment times. The dynamic-jaw mode should be used instead of the conventional fixed-jaw mode in tomotherapy for lung cancer.
© The Author(s) 2014.

Entities:  

Keywords:  Dynamic jaws; Lung cancer; Static port; Tomotherapy.

Mesh:

Year:  2014        PMID: 24945375      PMCID: PMC4639908          DOI: 10.7785/tcrtexpress.2013.600280

Source DB:  PubMed          Journal:  Technol Cancer Res Treat        ISSN: 1533-0338


Introduction

The TomoTherapy® (Accuray Inc., Sunnyvale, CA, USA) is a radiation delivery system that combines dynamic intensity-modulated radiation therapy (IMRT) and an on-board imaging system (1, 2). The role of tomotherapy has now been established for the treatment of various targets (3–14). With the conventional tomotherapy delivery mode, the cranio-caudal “penumbra”, i.e., dose scattering at the cranio-caudal edges of a target, has been an issue that should be improved. In the treatment of lung cancer with IMRT, low dose irradiation to the lung has been reported to be a risk factor for radiation pneumonitis (3, 4). To reduce the lung volume receiving low dose radiation, we reported the usefulness of tomotherapy using static ports (TomoDirect® mode) for the treatment of locally advanced lung cancer (5). However, scattered doses were not negligible, especially in the 5.0-cm fixed-jaw mode. Recently, a newly developed dynamic jaws technology (TomoEDGE®) has been introduced in our institution first in Japan. With this technology, radiation doses for the cranio-caudal edges of the target can be lowered by using narrower jaws around the edges (6–8). The purpose of this study is to evaluate the characteristics of dynamic-jaw-mode plans compared to conventional fixed-jaw-mode plans for lung cancer.

Methods

In 35 patients, four plans using the 2.5-cm dynamic-, 2.5-cm fixed-, 5.0-cm dynamic-, and 5.0-cm fixed-jaw modes were made and ompared. For 10 patients with upper lobe stage I lung cancer, helical tomotherapy plans were generated to obtain better conformity than with the TomoDirect mode. For 25 patients with locally advanced lung cancer, treatment plans using four ports with the TomoDirect mode were made to reduce the lung volume receiving low dose irradiation. The location of the primary tumor in the 25 patients was the upper lobe in 10, middle lobe in 2, lower lobe in 5, and mediastinum in 8. This is a planning comparison study, and actually the patients were treated with linac-based stereotactic radiotherapy for stage I lung cancer (15, 16) or conventional radiotherapy for locally advanced lung cancer.

CT Simulation and Planning

The 10 patients with stage I lung cancer were immobilized in a supine position with a vacuum bag system (BodyFIX; Medical Intelligence, Schwabmünchen, Germany) alongside the whole body. Computed tomography (CT) scans were performed with a slice thickness of 3.2-mm using a 4-row multi-detector CT (Mx8000; Philips Medical Systems, Best, The Netherlands) in a supine position under normal breathing, and with breath holding during the expiratory and inspiratory phases as described in detail previously (15). For the 25 patients with locally advanced lung cancer, actually treated with conventional radiotherapy, CT scans were performed under normal breathing without immobilization devices. CT images were reconstructed with a 2.5-mm thickness. Contouring of target volumes and normal structures was performed on the Pinnacle3 version 9 treatment planning system (Philips Medical System, Eindhoven, The Netherlands). The contours created in the treatment planning system were exported to the TomoTherapy treatment planning system (Tomo HD version 2.0), where all plans were generated. For stage I lung cancer, the clinical target volume (CTV) was defined as the visible gross tumor volume (GTV). The CTV on CT during the 3 phases were superimposed to represent the internal target volume. We defined the planning target volume (PTV) margin for the internal target volume as 5-mm in all directions. As a minimum coverage dose for 95% of the PTV (D95%), 56 Gy in 8 fractions was prescribed. We contoured the ispirateral lung and total lung excluding the PTV. As dose constraints, 1) D95% > 95% of the prescribed dose, 2) V90% of the PTV ≥ 95%, 3) mean lung dose (MLD) < 18 Gy, V20 Gy < 20% of the total lung, and 4) spinal cord maximum dose < 25 Gy were satisfied. V90% was defined as the percentage of the PTV receiving at least 90% of the prescribed dose. The Vx Gy value represents the percentage or absolute volume (V) receiving the specified dose (x) in Gy, e.g., V10 Gy is the percentage volume receiving 10 Gy. For the 25 patients with locally advanced lung cancer, contrast-enhanced CT images were acquired and fused to the planning CT images to delineate the target, but unenhanced CT images were used for dose calculation to keep calculation accuracy (17). The visible primary tumor, swollen lymph nodes, and prophylactic lymph node area were contoured as the CTV1, CTV2, and CTV3, respectively. For the PTV1, the CTV1 was expanded by 5-mm for mediastinal primary tumors, 10-mm in all directions for upper-lobe and 1 of 5 lower-lobe primary tumors that invaded the chest wall. For four patients with a lower-lobe primary tumor, we defined the margin as 15-mm in the cranio-caudal direction. The CTV2 and CTV3 were expanded by 5-mm in all directions for the PTV2 and PTV3, respectively. Using the simultaneous integrated boost technique, 60 Gy in 2-Gy daily fractions to the PTV1 and PTV2, and 51 Gy in 1.7-Gy fractions to the PTV3 were prescribed as median doses (18). We contoured the ispirateral lung and total lung excluding the PTV1. Dose constraints were: 1) D95% > 90% of the prescribed dose; 2) total lung: MLD > 17 Gy, V10 Gy < 40%, and V20 Gy < 30%; 3) spinal cord: maximum dose < 50 Gy; and 4) esophagus: maximum dose < 66 Gy. Inverse planning procedure of optimization using the TomoTherapy planning station was described in detail previously (5). The same pitch (0.287 or 0.215 for helical tomotherapy and 0.251 or 0.500 for TomoDirect) and modulation factor (2.0) were used in each patient. When the dose constrains could not be fulfilled, the dose of the cord and the esophagus took priority. A fine calculation grid (1.95 × 1.95-mm) was used for the final calculation process. To compare 2.5-cm dynamic-, 2.5-cm fixed-, 5.0-cm dynamic-, and 5.0-cm fixed-jaw-mode plans, the conformity index, uniformity index, dose distribution in organs at risk, and beam-on time were evaluated in the TomoTherapy planning system. The conformity index and uniformity index were calculated according to the following formulae (5, 19). where VPTV = PTV (ml), TVPV = lesion volume (ml) covered by the prescribed isodose, VTV = prescribed isodose volume (ml), and D5% = minimum dose delivered to the 5% of the PTV. The lower conformity index indicates the higher conformity, and the lower uniformity index indicates the better homogeneity. An ideal conformity index and uniformity index are both 1.

Statistical Analysis

The conformity index, uniformity index, dose distribution in organs at risk, integral dose, monitor unit, and beam-on time were compared. Paired t-test was used to analyze the difference within the same-size jaws as a priori comparisons. We used parametric analysis of variance for the dependent samples among the whole groups. These were performed when the normality of the distribution and homogeneity of the variance in the analyzed groups were confirmed by Kolmogorov-Sminov test and Bartlett’s test. When these criteria were not met, Wilcoxon signed-rank test as a priori comparisons and Friedman analysis of variance for the whole groups were performed. Statistical analyses were carried out with the statistical software package ‘R’ (20). All plannings and evaluations were performed by one radiation oncologist (Y. M.).

Results

Stage I Lung Cancer

Figure 1 shows representative dose distributions for the four plans in a patient with stage I lung cancer. The treatment parameters, dose-volume parameters, beam-on times, monitor units, and gantry periods of the four plans for 10 patients are summarized in Tables I and II. The conformity index of the PTV and the maximum dose of the esophagus were similar among the four plans. The uniformity index was slightly better in the dynamic-jaw-mode than fixed-jaw-mode plans, but these differences appeared to be clinically of no significance (Table I). The differences in the maximum spinal cord dose were small between the dynamic-jaw and fixed-jaw plans. All dose-volume parameters for the whole body (V5-30 Gy) and lung (V5-50 Gy and MLD) were better in the 2.5-cm dynamic-jaw plans than in the 2.5-cm fixed-jaw plans; this was also observed between the 5.0-cm dynamic-jaw and fixed-jaw plans, except V40 Gy for the ipsilateral lung and V50 Gy for the total and ipsilateral lungs (Tables I and II; Figure 2A and 2B). The V10 Gy of the contralateral lung were all <1% (data not shown). The mean of the V5 Gy of the total lung, ispilateral lung, and the whole body in 5.0-cm dynamic-jaw plans were lower than in 2.5-cm fixed-jaw plans (Figure 2A and 2B). The beam-on time was longer by 32-54 seconds in dynamic-jaw plans probably due to slower gantry rotation and higher monitor units. The 5.0-cm dynamic-jaw plans reduced the beam-on time by 21-27% compared to the 2.5-cm jaw plans (Figure 2C).
Figure 1:

Dose distribution of the four mode plans in a patient with stage I lung cancer. Cranio-caudal dose penumbra can be reduced using dynamic jaw mode (arrows).

Table I

Treatment and dose-volume parameters, monitor units, gantry period, and beam-on times of the four helical tomotherapy plans for patients with stage I lung cancer.

Mean ± standard deviation
2.5-cm Dynamic2.5-cm Fixedp*5.0-cm Dynamic5.0-cm Fixedp**p***
Patient number10
Total PTV (ml)22.7 ± 10.3
Pitch0.215, 0.287
Modulation factor2.0
Conformity index1.28 ± 0.121.30 ± 0.060.321.29 ± 0.050.160.39
Uniformity index1.09 ± 0.021.10 ± 0.020.02||1.10 ± 0.021.11 ± 0.030.05||0.0002§
Monitor unit5525 ± 6445066 ± 6350.0023988 ± 2453219 ± 3630.002<0.0001
Gantry period (sec)38.1 ± 5.835.2 ± 5.1<0.0001||41 ± 8.332.1 ± 3.90.002||<0.0001§
Beam-on time (sec)397.9 ± 45.1365.9 ± 44.60.002290.3 ± 17.4236.2 ± 25.40.002<0.0001
Body
 V5 Gy (ml)1825.6 ± 409.42143.4 ± 441.6<0.0001||1983.4 ± 401.62938.9 ± 559.9<0.0001||<0.0001§
 V10 Gy (ml)806.1 ± 255.2909.0 ± 271.8<0.0001||881.5 ± 264.71196.4 ± 335.7<0.0001||<0.0001§
 V20 Gy (ml)240.3 ± 89.5269.5 ± 91.3<0.0001||265.6 ± 96.3354.0 ± 113.2<0.0001||< 0.0001§
 V30 Gy (ml)115.4 ± 43.3128.0 ± 43.4<0.0001||127.9 ± 46.4163.4 ± 51.9<0.0001||< 0.0001§
Spinal cord maximum (Gy)10.3 ± 4.610.9 ± 4.80.003||9.2 ± 4.68.9 ± 4.90.80||0.02§
Esophagus maximum (Gy)14.1 ± 4.114.3 ± 4.20.14||14.0 ± 3.814.1 ± 4.10.68||0.31§

Abbreviation: PTV: Planning target volume.

*p-value between 2.5-cm dynamic- and 2.5-cm fixed- jaw plans.

**p-value between 5.0-cm dynamic- and 5.0-cm fixed- jaw plans.

***p-value among the four plans.

†Calculated by Friedman analysis of variance.

‡Calculated by Wilcoxon signed-rank test.

§Calculated by parametric analysis of variance for depended samples.

||Calculated by paired t-test.

Table II

Dose-volume parameters of the four helical tomotherapy plans for patients with stage I lung cancer.

Mean ± standard deviation
2.5-cm Dynamic2.5-cm Fixedp*5.0-cm Dynamic5.0-cm Fixedp**p***
Total lung
 V5 Gy (%)18.5 ± 6.821.8 ± 7.5<0.0001||20.1 ± 7.129.4 ± 9.80.0001||<0.0001§
 V10 Gy (%)9.2 ± 2.810.8 ± 3.2<0.0001||10.0 ± 2.914.5 ± 4.0<0.001||<0.0001§
 V20 Gy (%)4.0 ± 1.44.7 ± 1.6<0.0001||4.4 ± 1.56.2 ± 1.9<0.0001||<0.0001§
 V30 Gy (%)1.9 ± 0.92.3 ± 0.80.009||2.3 ± 0.83.0 ± 1.0<0.0001||<0.0001§
 V40 Gy (%)1.0 ± 0.41.2 ± 0.5<0.0001||1.2 ± 0.51.5 ± 0.5<0.0001||<0.0001§
 V50 Gy (%)0.4 ± 0.20.5 ± 0.20.01||0.5 ± 0.20.5 ± 0.20.06||<0.0001§
 MLD (Gy)3.3 ± 0.93.9 ± 1.0<0.0001||3.6 ± 0.95.1 ± 1.3<0.0001||<0.0001§
Ispilateral lung
 V5 Gy (%)25.3 ± 7.131.1 ± 8.4<0.0001||27.0 ± 7.442.8 ± 10.6<0.0001||<0.0001§
 V10 Gy (%)18.0 ± 5.821.3 ± 6.5<0.0001||19.5 ± 5.928.4 ± 7.7<0.0001||<0.0001§
 V20 Gy (%)8.1 ± 3.69.4 ± 4.0<0.0001||8.9 ± 3.712.4 ± 4.5<0.0001||<0.0001§
 V30 Gy (%)4.0 ± 1.94.7 ± 2.1<0.0001||4.6 ± 2.06.2 ± 2.3<0.0001||<0.0001§
 V40 Gy (%)2.1 ± 1.12.4 ± 1.1<0.0001||2.4 ± 1.13.7 ± 2.20.09||0.009§
 V50 Gy (%)0.9 ± 0.50.9 ± 0.50.031.0 ± 0.61.6 ± 1.60.06<0.0001
Contralateral lung
V5 Gy (%)11.6 ± 7.112.5 ± 7.10.007||13.1 ± 7.515.7 ± 9.20.004||<0.0001§

*p-value between 2.5-cm dynamic- and 2.5-cm fixed-jaw plans.

**p-value between 5.0-cm dynamic- and 5.0-cm fixed-jaw plans.

***p-value among the four plans.

†Calculated by Friedman analysis of variance.

‡Calculated by Wilcoxon signed-rank test.

§Calculated by parametric analysis of variance for depended samples.

||Calculated by paired t-test.

Figure 2:

V5 Gy of the total lung (A), V5 Gy of the whole body (B), and beam-on time (C) in each mode for patients with stage I lung cancer. The box includes the central 50% of data (25-75%), and the central 99% of data are contained within the error bars. The solid line within each box indicates the median of the data. V5 Gy refers to the percentage or absolute volume receiving 5 Gy, 2.5D refers to 2.5-cm dynamic-jaw mode, 2.5F refers to 2.5-cm fixed-jaw mode, 5.0D refers to 5.0-cm dynamic-jaw mode, and 5.0F refers to 5.0-cm fixed-jaw mode. *p < 0.01.

Dose distribution of the four mode plans in a patient with stage I lung cancer. Cranio-caudal dose penumbra can be reduced using dynamic jaw mode (arrows). Treatment and dose-volume parameters, monitor units, gantry period, and beam-on times of the four helical tomotherapy plans for patients with stage I lung cancer. Abbreviation: PTV: Planning target volume. *p-value between 2.5-cm dynamic- and 2.5-cm fixed- jaw plans. **p-value between 5.0-cm dynamic- and 5.0-cm fixed- jaw plans. ***p-value among the four plans. †Calculated by Friedman analysis of variance. ‡Calculated by Wilcoxon signed-rank test. §Calculated by parametric analysis of variance for depended samples. ||Calculated by paired t-test. Dose-volume parameters of the four helical tomotherapy plans for patients with stage I lung cancer. *p-value between 2.5-cm dynamic- and 2.5-cm fixed-jaw plans. **p-value between 5.0-cm dynamic- and 5.0-cm fixed-jaw plans. ***p-value among the four plans. †Calculated by Friedman analysis of variance. ‡Calculated by Wilcoxon signed-rank test. §Calculated by parametric analysis of variance for depended samples. ||Calculated by paired t-test. V5 Gy of the total lung (A), V5 Gy of the whole body (B), and beam-on time (C) in each mode for patients with stage I lung cancer. The box includes the central 50% of data (25-75%), and the central 99% of data are contained within the error bars. The solid line within each box indicates the median of the data. V5 Gy refers to the percentage or absolute volume receiving 5 Gy, 2.5D refers to 2.5-cm dynamic-jaw mode, 2.5F refers to 2.5-cm fixed-jaw mode, 5.0D refers to 5.0-cm dynamic-jaw mode, and 5.0F refers to 5.0-cm fixed-jaw mode. *p < 0.01.

Locally Advanced Lung Cancer

Figure 3 shows representative dose distributions for the four plans in a patient with locally advanced lung cancer. The treatment parameters, dose-volume parameters, beam-on times, and monitor units of the four plans for all 25 patients are summarized in Tables III and IV. The conformity index and uniformity index of the PTV and the maximum spinal cord dose were similar when all four plans were compared (Table III). The difference in the maximum esophagus dose did not seem to be of clinical significance. The V40 Gy for the heart was slightly lower in the dynamic-jaw than fixed-jaw plans. The V5-30 Gy for the whole body, MLD and V5-20 Gy for the total, ipsilateral, and contralateral lungs were consistently lower in the dynamic-jaw than fixed-jaw plans (Tables III and IV; Figure 4A and 4B). For V30-50 Gy, the superiority of either plan was not constant. The beam-on time was longer in dynamic-jaw plans probably due to higher monitor units, but the difference was relatively small (about 7 seconds longer in the dynamic-jaw plans). The 5.0-cm dynamic-jaw plans reduced the beam-on time by 38-39% compared to the 2.5-cm jaw plans (Figure 4C).
Figure 3:

Dose distribution of the four mode plans in a patient with locally advanced lung cancer. Cranio-caudal dose penumbra can be reduced using dynamic jaw mode (arrows).

Table III

Treatment and dose-volume parameters, monitor units, and beam-on times of the four TomoDirect plans for patients with locally advanced lung cancer.

Mean ± standard deviation
2.5-cm Dynamic2.5-cm Fixedp*5.0-cm Dynamic5.0-cm Fixedp**p***
Patient number25
Total PTV (ml)338.7 ± 179.0
Pitch0.2510.500
Modulation factor2.0
Conformity index12.5 ± 13.013.3 ± 13.80.00211.3 ± 10.413.3 ± 12.30.770.08
Uniformity index1.09 ± 0.041.10 ± 0.050.003||1.10 ± 0.051.11 ± 0.050.0001||0.08§
Monitor unit3509 ± 6013417 ± 596<0.00011912 ± 2941806 ± 292<0.0001<0.0001
Beam-on time (sec)287.9 ± 42.1281.3 ± 41.9<0.0001175.7 ± 20.6168.3 ± 20.4<0.0001<0.0001
Body
 V5 Gy (ml)4848.5 ± 1337.55197.4 ± 1402.4<0.0001||5220.2 ± 1442.86153.9 ± 1599.0<0.0001||<0.0001§
 V10 Gy (ml)4287.4 ± 1175.34602.2 ± 1241.8<0.0001||4628.7 ± 1282.55418.1 ± 1433.5<0.0001||<0.0001§
 V20 Gy (ml)3737.7 ± 1046.83929.5 ± 1089.1<0.0001||3999.5 ± 1134.94500.7 ± 1247.8<0.0001||<0.0001§
 V30 Gy (ml)3282.8 ± 933.63405.4 ± 972.5<0.0001||3488.7 ± 1000.23762.4 ± 1074.1<0.0001||<0.0001§
Spinal cord maximum (Gy)43.3 ± 2.243.5 ± 2.40.11||43.1 ± 2.243.1 ± 2.00.94||0.09§
Esophagus maximum (Gy)59.4 ± 2.259.0 ± 2.40.05||58.2 ± 2.758.3 ± 2.80.27||<0.0001§
Heart V40 Gy (%)7.9 ± 7.78.3 ± 7.70.01||10.4 ± 9.611.2 ± 9.60.02||<0.0001§

Abbreviation: PTV: Planning target volume.

*p-value between 2.5-cm dynamic- and 2.5-cm fixed-jaw plans.

**p-value between 5.0-cm dynamic- and 5.0-cm fixed-jaw plans.

***p-value among the four plans.

†Calculated by Friedman analysis of variance.

‡Calculated by Wilcoxon signed-rank test.

§Calculated by parametric analysis of variance for depended samples.

||Calculated by paired t-test.

Table IV

Dose-volume parameters of the four TomoDirect plans for patients with locally advanced lung cancer.

Mean ± standard deviation
2.5-cm Dynamic2.5-cm Fixedp*5.0-cm Dynamic5.0-cm Fixedp**p***
Total lung
 V5 Gy (%)37.3 ± 11.038.7 ± 11.4<0.0001||40.4 ± 11.744.0 ± 12.4<0.0001||<0.0001§
 V10 Gy (%)31.1 ± 9.632.3 ± 10.0<0.0001||34.4 ± 10.637.1 ± 11.2<0.0001||<0.0001§
 V20 Gy (%)25.0 ± 7.925.5 ± 8.40.007||27.5 ± 8.929.2 ± 9.5<0.0001||<0.0001§
 V30 Gy (%)20.7 ± 6.621.0 ± 7.00.09||22.7 ± 7.423.3 ± 7.7<0.0001||<0.0001§
 V40 Gy (%)17.2 ± 6.216.8 ± 5.80.40||18.1 ± 6.118.1 ± 6.20.82||<0.0001§
 V50 Gy (%)12.7 ± 5.312.3 ± 4.20.43||12.8 ± 4.612.5 ± 4.60.0006||0.53§
MLD (Gy)13.7 ± 4.014.1 ± 4.1<0.0001||14.8 ± 4.415.6 ± 4.5<0.0001||<0.0001§
Ispilateral lung
 V5 Gy (%)53.5 ± 19.455.5 ± 20.0<0.0001||56.8 ± 19.762.7 ± 20.7<0.0001||<0.0001§
 V10 Gy (%)46.3 ± 17.448.2 ± 18.1<0.0001||50.0 ± 18.155.2 ± 19.6<0.0001||<0.0001§
 V20 Gy (%)38.4 ± 14.739.6 ± 15.1<0.0001||42.4 ± 16.045.3 ± 17.1<0.0001||<0.0001§
 V30 Gy (%)32.6 ± 12.933.3 ± 13.10.0002||35.9 ± 14.137.0 ± 14.4<0.0001||<0.0001§
 V40 Gy (%)26.9 ± 11.227.3 ± 11.30.03||29.3 ± 12.228.8 ± 2.20.22||0.009§
 V50 Gy (%)20.6 ± 9.121.0 ± 9.40.13||21.7 ± 9.821.1 ± 9.30.003||0.003§
Contralateral lung
 V5 Gy (%)23.4 ± 9.924.1 ± 10.3<0.0001||27.0 ± 11.928.7 ± 13.00.0005||<0.0001§
 V10 Gy (%)18.1 ± 8.118.8 ± 8.40.0008||21.1 ± 9.922.2 ± 10.90.002||<0.0001§
 V20 Gy (%)13.4 ± 6.813.6 ± 7.00.006||15.3 ± 8.215.8 ± 8.70.007||<0.0001§
 V30 Gy (%)10.6 ± 6.010.6 ± 6.10.49||11.7 ± 7.112.0 ± 7.30.02||0.001§
 V40 Gy (%)8.2 ± 5.28.2 ± 5.20.41||8.9 ± 5.99.0 ± 6.00.12||0.03§
 V50 Gy (%)5.2 ± 3.95.2 ± 3.90.65||5.6 ± 4.55.6 ± 4.40.66||0.29§

*p-value between 2.5-cm dynamic- and 2.5-cm fixed-jaw plans.

**p-value between 5.0-cm dynamic- and 5.0-cm fixed-jaw plans.

***p-value among the four plans.

§Calculated by parametric analysis of variance for depended samples.

||Calculated by paired t-test.

Figure 4:

V5 Gy of the total lung (A), V5 Gy of the whole body (B), and beam-on time (C) in each mode for patients with locally advanced lung cancer. The box includes the central 50% of data (25-75%), and the central 99% of data are contained within the error bars. The solid line within each box indicates the median of the data. V5 Gy refers to the percentage or absolute volume receiving 5 Gy, 2.5D refers to 2.5-cm dynamic-jaw mode, and 2.5F refers to 2.5-cm fixed-jaw mode, 5.0D refers to 5.0-cm dynamic-jaw mode, and 5.0F refers to 5.0-cm fixed-jaw mode. $p < 0.01.

Dose distribution of the four mode plans in a patient with locally advanced lung cancer. Cranio-caudal dose penumbra can be reduced using dynamic jaw mode (arrows). Treatment and dose-volume parameters, monitor units, and beam-on times of the four TomoDirect plans for patients with locally advanced lung cancer. Abbreviation: PTV: Planning target volume. *p-value between 2.5-cm dynamic- and 2.5-cm fixed-jaw plans. **p-value between 5.0-cm dynamic- and 5.0-cm fixed-jaw plans. ***p-value among the four plans. †Calculated by Friedman analysis of variance. ‡Calculated by Wilcoxon signed-rank test. §Calculated by parametric analysis of variance for depended samples. ||Calculated by paired t-test. Dose-volume parameters of the four TomoDirect plans for patients with locally advanced lung cancer. *p-value between 2.5-cm dynamic- and 2.5-cm fixed-jaw plans. **p-value between 5.0-cm dynamic- and 5.0-cm fixed-jaw plans. ***p-value among the four plans. §Calculated by parametric analysis of variance for depended samples. ||Calculated by paired t-test. V5 Gy of the total lung (A), V5 Gy of the whole body (B), and beam-on time (C) in each mode for patients with locally advanced lung cancer. The box includes the central 50% of data (25-75%), and the central 99% of data are contained within the error bars. The solid line within each box indicates the median of the data. V5 Gy refers to the percentage or absolute volume receiving 5 Gy, 2.5D refers to 2.5-cm dynamic-jaw mode, and 2.5F refers to 2.5-cm fixed-jaw mode, 5.0D refers to 5.0-cm dynamic-jaw mode, and 5.0F refers to 5.0-cm fixed-jaw mode. $p < 0.01.

Discussion

It is well known that V20 Gy of the lung and the MLD are predictors of pulmonary complications (21–23). In addition, low dose irradiation to the lung as expressed by V5 Gy or V10 Gy has also been reported to be a risk factor for pulmonary toxicity in IMRT of lung cancer, especially when combined with chemotherapy (3, 4). This study suggested that low dose irradiation to the lung could be reduced by using the dynamic-jaw mode in tomotherapy for lung cancer. For the patients with stage I lung cancer, although the beam-on time was relatively long when the dynamic-jaw mode was used, dose-volume parameters of plans using the dynamic-jaw mode were better than those using the fixed-jaw mode. The 2.5-cm dynamic-jaw mode appeared to be the best among these four modes in terms of dose distribution. It is notable that those parameters including the integral dose in the 5.0-cm dynamic-jaw mode were superior to those in the 2.5-cm conventional fixed jaw mode while reducing the beam-on time by 21%. Thus, 5.0-cm dynamic-jaw mode would be suitable for patients who have difficulty in lying immobile during a longer treatment time. For the patients with locally advanced lung cancer, dose-volume parameters of plans using the dynamic-jaw mode were generally better than those using the fixed-jaw mode, and the differences in beam-on time were small between the plans; therefore, the dynamic-jaw mode is generally superior to the conventional fixed-jaw mode for the treatment of locally advanced lung cancer. The 2.5-cm dynamic-jaw mode appeared to be the best among these four modes in terms of dose distribution. The 5.0-cm dynamic-jaw mode required shorter treatment time while maintaining dose distribution comparable to that obtained by the conventional 2.5-cm fixed-jaw mode. Thus, this mode might be alternatively used for patients for whom a shorter treatment time is desirable as well as for patients with stage I lung cancer. In this planning comparison study, CT simulation data without immobilization devices was used for the patients with locally advanced lung cancer. However, immobilization devices should be used in clinical practice when using IMRT technique (24). A few planning comparison studies using the dynamic-jaw and dynamic-couch modes have been reported for tumors other than lung cancer (6–8). This mode offers dynamic jaw alignment throughout the treatment and dynamic couch speed, but it is not available for clinical application yet. The previous studies only employed the TomoHelical mode. With these new options of tomotherapy, beam-on time could be reduced by 66% compared to the conventional 2.5-cm fixed-jaw mode (from 595 to 199 seconds) in the treatment of nasopharyngeal cancer without significant difference in dose distribution (6). The current study showed that the dynamic-jaw mode is useful when used with both TomoHelical and TomoDirect modes. In the present study, the 5.0-cm dynamic-jaw mode could reduce beam-on time by 21-39% in the treatment of lung cancer. Adding the dynamic couch mode may achieve a shorter treatment time, but the characteristics of dose distribution including those in the lung and the heart are unclear for the lung target volume. Further investigation is warranted.

Conclusion

The dynamic-jaw mode should be used instead of the conventional fixed-jaw mode in tomotherapy for lung cancer. The 2.5-cm dynamic-jaw mode would be the best in terms of dose distribution. The 5.0-cm dynamic-jaw mode might be an alternative to reduce beam-on time.
  23 in total

1.  [The ICRU Report 83: prescribing, recording and reporting photon-beam intensity-modulated radiation therapy (IMRT)].

Authors:  N Hodapp
Journal:  Strahlenther Onkol       Date:  2012-01       Impact factor: 3.621

2.  Intensity-modulated adjuvant whole breast radiation delivered with static angle tomotherapy (TomoDirect): a prospective case series.

Authors:  Pierfrancesco Franco; Michele Zeverino; Fernanda Migliaccio; Piera Sciacero; Domenico Cante; Valeria Casanova Borca; Paolo Torielli; Cecilia Arrichiello; Giuseppe Girelli; Gianmauro Numico; Maria Rosa La Porta; Santi Tofani; Umberto Ricardi
Journal:  J Cancer Res Clin Oncol       Date:  2013-09-14       Impact factor: 4.553

3.  Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy.

Authors:  T R Mackie; T Holmes; S Swerdloff; P Reckwerdt; J O Deasy; J Yang; B Paliwal; T Kinsella
Journal:  Med Phys       Date:  1993 Nov-Dec       Impact factor: 4.071

4.  Treatment-related pneumonitis and acute esophagitis in non-small-cell lung cancer patients treated with chemotherapy and helical tomotherapy.

Authors:  Chang Hoon Song; Hongryull Pyo; Sung Ho Moon; Tae Hyun Kim; Dae Woong Kim; Kwan Ho Cho
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-06       Impact factor: 7.038

5.  Prediction of radiation pneumonitis: dose-volume histogram analysis in 62 patients with non-small cell lung cancer after three-dimensional conformal radiotherapy.

Authors:  T Piotrowski; M Matecka-Nowak; P Milecki
Journal:  Neoplasma       Date:  2005       Impact factor: 2.575

6.  Impact of toxicity grade and scoring system on the relationship between mean lung dose and risk of radiation pneumonitis in a large cohort of patients with non-small cell lung cancer.

Authors:  Susan L Tucker; Hekun Jin; Xiong Wei; Shulian Wang; Mary K Martel; Ritsuko Komaki; H Helen Liu; Radhe Mohan; Yuhchyau Chen; James D Cox; Zhongxing Liao
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-10-14       Impact factor: 7.038

7.  Tomotherapy radiosurgery for arteriovenous malformations--current possibilities and future options with helical tomotherapy dynamic jaws?

Authors:  S Krause; S Beck; O Schramm; K Schubert; H Hauswald; A Zabel-du Bois; K Herfarth; J Debus; F Sterzing
Journal:  Technol Cancer Res Treat       Date:  2013-03-26

8.  Predictive factors for radiation pneumonitis in lung cancer treated with helical tomotherapy.

Authors:  Youngkyong Kim; Seong Eon Hong; Moonkyoo Kong; Jinhyun Choi
Journal:  Cancer Res Treat       Date:  2013-12-31       Impact factor: 4.679

9.  Outcome and toxicity of stereotactic body radiotherapy with helical tomotherapy for inoperable lung tumor: analysis of Grade 5 radiation pneumonitis.

Authors:  Norihiro Aibe; Hideya Yamazaki; Satoaki Nakamura; Takuji Tsubokura; Kana Kobayashi; Naohiro Kodani; Takuya Nishimura; Haruumi Okabe; Kei Yamada
Journal:  J Radiat Res       Date:  2014-01-23       Impact factor: 2.724

10.  Stereotactic body radiotherapy for stage I lung cancer and small lung metastasis: evaluation of an immobilization system for suppression of respiratory tumor movement and preliminary results.

Authors:  Fumiya Baba; Yuta Shibamoto; Natsuo Tomita; Chisa Ikeya-Hashizume; Kyota Oda; Shiho Ayakawa; Hiroyuki Ogino; Chikao Sugie
Journal:  Radiat Oncol       Date:  2009-05-28       Impact factor: 3.481

View more
  13 in total

1.  Chemoradiotherapy for localized extranodal natural killer/T-cell lymphoma, nasal type, using a shrinking-field radiation strategy: multi-institutional experience.

Authors:  Yukiko Hattori; Taro Murai; Hiromitsu Iwata; Kaoru Uchiyama; Mikio Mimura; Eriko Kato; Rumi Murata; Yuta Shibamoto
Journal:  Jpn J Radiol       Date:  2016-02-06       Impact factor: 2.374

2.  Survey of malignant pleural mesothelioma treatment in Japan: Patterns of practice and clinical outcomes in tomotherapy facilities.

Authors:  Mikiko Nakanishi-Imai; Taro Murai; Masahiro Onishi; Atsuto Mouri; Takafumi Komiyama; Motoko Omura; Shigehiro Kudo; Akihiko Miyamoto; Masaru Hoshino; Shinichi Ogawa; Shizuko Ohashi; Masahiko Koizumi; Junichi Omagari; Hiroshi Mayahara; Katsuyuki Karasawa; Toshiyuki Okumura; Yuta Shibamoto
Journal:  J Radiat Res       Date:  2022-03-17       Impact factor: 2.724

3.  Dose-volume histogram comparison between static 5-field IMRT with 18-MV X-rays and helical tomotherapy with 6-MV X-rays.

Authors:  Akihiro Hayashi; Yuta Shibamoto; Yukiko Hattori; Takeshi Tamura; Michio Iwabuchi; Shinya Otsuka; Chikao Sugie; Takeshi Yanagi
Journal:  J Radiat Res       Date:  2015-01-20       Impact factor: 2.724

4.  Efficacy of the Dynamic Jaw Mode in Helical Tomotherapy With Static Ports for Breast Cancer.

Authors:  Chikao Sugie; Yoshihiko Manabe; Akihiro Hayashi; Taro Murai; Taiki Takaoka; Yukiko Hattori; Hiromitsu Iwata; Ran Takenaka; Yuta Shibamoto
Journal:  Technol Cancer Res Treat       Date:  2014-11-14

5.  Efficacy of stereotactic radiotherapy for brain metastases using dynamic jaws technology in the helical tomotherapy system.

Authors:  Taro Murai; Akihiro Hayashi; Yoshihiko Manabe; Chikao Sugie; Taiki Takaoka; Takeshi Yanagi; Tetsuya Oguri; Masayuki Matsuo; Yoshimasa Mori; Yuta Shibamoto
Journal:  Br J Radiol       Date:  2016-08-24       Impact factor: 3.039

6.  A retrospective comparison of outcome and toxicity of preoperative image-guided intensity-modulated radiotherapy versus conventional pelvic radiotherapy for locally advanced rectal carcinoma.

Authors:  Chun-Ming Huang; Ming-Yii Huang; Hsiang-Lin Tsai; Ching-Wen Huang; Cheng-Jen Ma; Chih-Hung Lin; Chih-Jen Huang; Jaw-Yuan Wang
Journal:  J Radiat Res       Date:  2017-03-01       Impact factor: 2.724

7.  Intensity-modulated radiation therapy for multiple targets with tomotherapy using multiple sets of static ports from different angles.

Authors:  Yoshihiko Manabe; Yuta Shibamoto; Taro Murai; Akira Torii; Masanari Niwa; Takuhito Kondo; Dai Okazaki; Chikao Sugie
Journal:  J Appl Clin Med Phys       Date:  2020-04-06       Impact factor: 2.102

8.  Dosimetric comparison of helical tomotherapy and conventional Linac-based X-knife stereotactic body radiation therapy for primary lung cancer or pulmonary metastases.

Authors:  Shuangshuang Li; Ju Yang; Juan Liu; Shanbao Gao; Baorui Liu; Jing Yan
Journal:  J Thorac Dis       Date:  2018-02       Impact factor: 2.895

9.  Definitive Radiotherapy With SBRT or IMRT Boost for Breast Cancer: Excellent Local Control and Cosmetic Outcome.

Authors:  Yuta Shibamoto; Taro Murai; Kazushi Suzuki; Chisa Hashizume; Kengo Ohta; Yuki Yamada; Masanari Niwa; Akira Torii; Masashi Shimohira
Journal:  Technol Cancer Res Treat       Date:  2018-01-01

10.  Feasibility of a Skin Dose Reduction for Nasopharyngeal Carcinoma Treated With High-Intensity-Modulated Delivery Techniques.

Authors:  Xiongfei Liao; Jie Li; Pei Wang; Xinghong Yao; Yulei Zhang; Tingqiang Tan; Lucia Clara Orlandini
Journal:  Technol Cancer Res Treat       Date:  2018-01-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.