Literature DB >> 17523041

Screening for refractive errors in preschool children with the vision screener.

Oliver Ehrt1, Anne Weber, Klaus-Peter Boergen.   

Abstract

BACKGROUND: The Vision Screener is a new, commercial version of the Power Refractor, an off-axis, hand-held video refractor to screen for amblyogenic refractive errors. The aim of our study was to determine the reproducibility of the measurements, compare them to cycloplegic refraction, and evaluate the sensitivity and specificity for the detection of amblyogenic refractive errors. PATIENTS AND METHODS: Included in the study were 161 preschool children, age 0.5-7.2 years without manifest strabismus > 10 degrees , who attended the outpatient clinic. After three measurements with the Vision Screener (version 4.3.15) and one measurement with +3 dpt glasses, the child underwent cycloplegic refraction.
RESULTS: The reproducibility of the three non-cycloplegic measurements was +/- 0.5 dpt in 85% and +/- 1 dpt in more than 95% of the children. Accommodation reduced the manifest hyperopia that could be measured with the Vision Screener by up to 3 dpt in some children. Compliance with the +3 dpt glasses increased from 20% in those under one year of age to > 75 % in those over three. The glasses did increase the sensitivity from 70% to 80% but decreased the specificity from 80% to 65%.
CONCLUSION: The Vision Screener is easy to handle and gives reproducible results. The sensitivity and specificity are between 70 and 80%. The +3 dpt glasses are only helpful in a few patients. The latest software version with a new algorithm is now under evaluation.

Entities:  

Mesh:

Year:  2007        PMID: 17523041     DOI: 10.1080/09273970601174969

Source DB:  PubMed          Journal:  Strabismus        ISSN: 0927-3972


  3 in total

1.  Two-dimensional simulation of eccentric photorefraction images for ametropes: factors influencing the measurement.

Authors:  Yifei Wu; Larry N Thibos; T Rowan Candy
Journal:  Ophthalmic Physiol Opt       Date:  2018-05-07       Impact factor: 3.117

2.  [Possibilities and limitations of amblyopia screening with auto-refractometers].

Authors:  O Ehrt
Journal:  Ophthalmologe       Date:  2016-04       Impact factor: 1.059

Review 3.  Plusoptix photoscreener use for paediatric vision screening in Flanders and Iran.

Authors:  Parinaz Rostamzad; Anna M Horwood; Nicoline E Schalij-Delfos; Kristel Boelaert; Harry J de Koning; Huibert J Simonsz
Journal:  Acta Ophthalmol       Date:  2019-06-14       Impact factor: 3.761