[Rebound tonometry in a clinical setting. Comparison with applanation tonometry].

C López-Caballero1, I Contreras, F J Muñoz-Negrete, G Rebolleda, L Cabrejas, P Marcelo.   

Abstract

PURPOSE: Rebound tonometry has recently been modified for its use in humans. The purpose of our study was to determine the precision of the ICare(R) rebound tonometer (RBT) as compared with the Goldmann applanation tonometer (GAT).
METHODS: Patients were recruited from our Hospital's Glaucoma Unit. In each patient, intraocular pressure (IOP) was measured without anaesthesia using the RBT and ten minutes later using the GAT. Central corneal thickness (CCT) was measured by pachymetry.
RESULTS: Sixty-eight patients were recruited (132 eyes). Mean IOP readings with RBT were 18.9 +/- 7.2 (SD) mmHg and were 15.5 +/- 5.7 mmHg with GAT (p<0.001). There was a good correlation between both instruments (r = 0.87, p<0.001). In most cases (84.6%) the IOP measured with the RBT was greater than that measured with the GAT. The mean difference between both instruments was 3.4 +/- 3.6 mmHg. There was a trend towards greater differences between those obtained using the RBT than the GAT when IOP values were higher. A statistically significant correlation was found between IOP readings with both tonometers and CCT, with higher differences being seen as the CCT increased.
CONCLUSION: The RBT can be employed in a clinical setting taking into account that it usually overestimates IOP as compared with the GAT. It could be especially useful in glaucoma screening campaigns since it can be operated by a trained technician.

Entities:  

Mesh:

Year:  2007        PMID: 17516263     DOI: 10.4321/s0365-66912007000500005

Source DB:  PubMed          Journal:  Arch Soc Esp Oftalmol        ISSN: 0365-6691


  4 in total

1.  Early and frequent development of ocular hypertension in children with nephrotic syndrome.

Authors:  Emi Kawaguchi; Kenji Ishikura; Riku Hamada; Yoshinobu Nagaoka; Yoshihiko Morikawa; Tomoyuki Sakai; Yuko Hamasaki; Hiroshi Hataya; Eiichiro Noda; Masaru Miura; Takashi Ando; Masataka Honda
Journal:  Pediatr Nephrol       Date:  2014-05-24       Impact factor: 3.714

2.  The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Authors:  Teresio Avitabile; Antonio Longo; Daniela Rocca; Roberta Amato; Caterina Gagliano; Marine Castaing
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-29       Impact factor: 3.117

3.  Effect of 0.5 mm larger donor corneal size on intraocular pressure following penetrating keratoplasty.

Authors:  Sagarika Patyal; Ajay Banarji; V S Gurunadh; Madhu Bhadauria; Poninder Kumar; A K Upadhyay
Journal:  Med J Armed Forces India       Date:  2011-07-21

4.  Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients.

Authors:  Shaoying Tan; Marco Yu; Nafees Baig; Linda Hansapinyo; Clement C Tham
Journal:  Sci Rep       Date:  2017-02-06       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.