Validation of the Microlife WatchBP Home device for self home blood pressure measurement according to the International Protocol.

George S Stergiou1, Periklis P Giovas, Charilaos P Gkinos, John D Patouras.   

Abstract

OBJECTIVE: Current guidelines recommend that self monitoring of blood pressure at home should only be performed using validated devices. This study assessed the accuracy of the Microlife WatchBP Home device for self home blood pressure measurement according to the European Society of Hypertension International Protocol.
METHODS: Thirty-three participants were included (15 in phase 1 and an additional 18 in phase 2). Simultaneous blood pressure measurements were taken by two observers (Y-tube-connected mercury sphygmomanometers) four times sequentially, with three measurements taken using the tested device. Absolute differences between observer and device measurements were classified into three zones (within 5, 10 and 15 mmHg). The number of measurements with a difference within 5 mmHg was calculated for each individual.
RESULTS: In phase 1, the device produced 38, 43 and 43 measurements within 5, 10 and 15 mmHg, respectively, for systolic blood pressure and 35, 45 and 45 for diastolic blood pressure. In phase 2.1, the device produced 75, 91 and 97 measurements within 5, 10 and 15 mmHg for systolic, and 74, 93 and 99 for diastolic blood pressure. In phase 2.2, 30 participants had at least two of their differences within 5 mmHg and two participants had no differences within 5 mmHg for systolic blood pressure, whereas for diastolic blood pressure the number of participants were 27 and three, respectively. Mean difference for systolic blood pressure was -0.3+/-5.6 mmHg and for diastolic -2.4+/-4.8 mmHg.
CONCLUSIONS: The Microlife WatchBP Home device for self home blood pressure measurement fulfills all the validation criteria of the International Protocol and can, therefore, be recommended for clinical use in the adult population.

Entities:  

Mesh:

Year:  2007        PMID: 17496469     DOI: 10.1097/MBP.0b013e3280b083ce

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  26 in total

1.  Automatic office blood pressure measured without doctors or nurses present.

Authors:  Joji Ishikawa; Efthimia G Nasothimiou; Nikos Karpettas; Scott McDoniel; Seth D Feltheimer; George S Stergiou; Thomas G Pickering; Joseph E Schwartz
Journal:  Blood Press Monit       Date:  2012-06       Impact factor: 1.444

2.  Patients' preference for ambulatory versus home blood pressure monitoring.

Authors:  E G Nasothimiou; N Karpettas; M G Dafni; G S Stergiou
Journal:  J Hum Hypertens       Date:  2013-10-24       Impact factor: 3.012

3.  Regular Acetaminophen Use and Blood Pressure in People With Hypertension: The PATH-BP Trial.

Authors:  Iain M MacIntyre; Emma J Turtle; Tariq E Farrah; Catriona Graham; James W Dear; David J Webb
Journal:  Circulation       Date:  2022-02-07       Impact factor: 39.918

4.  Association of Carbon Monoxide exposure with blood pressure among pregnant women in rural Ghana: Evidence from GRAPHS.

Authors:  Ashlinn K Quinn; Kenneth Ayuurebobi Ae-Ngibise; Darby W Jack; Ellen Abrafi Boamah; Yeetey Enuameh; Mohammed Nuhu Mujtaba; Steven N Chillrud; Blair J Wylie; Seth Owusu-Agyei; Patrick L Kinney; Kwaku Poku Asante
Journal:  Int J Hyg Environ Health       Date:  2015-11-10       Impact factor: 5.840

5.  Relationship between blood pressure variability and renal activity of the renin-angiotensin system.

Authors:  N Ozkayar; F Dede; F Akyel; T Yildirim; I Ateş; T Turhan; B Altun
Journal:  J Hum Hypertens       Date:  2015-07-30       Impact factor: 3.012

6.  Seasonal variation in meteorological parameters and office, ambulatory and home blood pressure: predicting factors and clinical implications.

Authors:  George S Stergiou; Aikaterini Myrsilidi; Anastasios Kollias; Antonios Destounis; Leonidas Roussias; Petros Kalogeropoulos
Journal:  Hypertens Res       Date:  2015-09-03       Impact factor: 3.872

7.  Excessive variability in systolic blood pressure that is self-measured at home exacerbates the progression of brain white matter lesions and cognitive impairment in the oldest old.

Authors:  Zhendong Liu; Yingxin Zhao; Hua Zhang; Qiang Chai; Yi Cui; Yutao Diao; Jianchao Xiu; Xiaolin Sun; Guosheng Jiang
Journal:  Hypertens Res       Date:  2015-12-03       Impact factor: 3.872

8.  Home-measured orthostatic hypotension associated with cerebral small vessel disease in a community-based older population.

Authors:  Yi Cui; Hua Zhang; Yingxin Zhao; Shangwen Sun; Qiang Chai; Gary Gong; Zhendong Liu
Journal:  Hypertens Res       Date:  2020-03-18       Impact factor: 3.872

9.  Validation of the Oregon Scientific BPU 330 for self-monitoring of blood pressure according to the International Protocol.

Authors:  Li Li; XinYu Zhang; ChunHong Yan; QingXiang Liang
Journal:  Vasc Health Risk Manag       Date:  2008

10.  Automated office blood pressure is in agreement with awake and mean 24-hour ambulatory blood pressure at the lower blood pressure range.

Authors:  Emmanuel A Andreadis; Charalampia V Geladari; Epameinondas T Angelopoulos
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-07-09       Impact factor: 3.738

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.