Validation of the Artsana CSI 610 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension.

Claudio Pini1, Marco Pastori, Jordan Baccheschi, Stefano Omboni, Gianfranco Parati.   

Abstract

BACKGROUND: There is evidence that blood pressure measurement outside the doctor's office can provide valuable information for the diagnostic evaluation of hypertensive patients and for monitoring their response to treatment. Home blood pressure monitoring devices have a major role in this setting, provided that their accuracy in measuring blood pressure is demonstrated by validation studies.
OBJECTIVE: This study aimed at verifying whether the automatic electronic oscillometric blood pressure measuring device Artsana CSI 610 complied with the standard of accuracy indicated by the ESH International Protocol.
METHODS: Sequential measurements of systolic and diastolic blood pressure were obtained in 33 participants using the mercury sphygmomanometer (two observers) and the test device (one supervisor). A standard adult cuff was always employed during the study. According to the ESH validation protocol, 99 couples of test device and reference blood pressure measurements were obtained during the two phases of the study (three pairs for each of the 33 participants).
RESULTS: The Artsana CSI 610 device successfully passed phase 1 of study validation with the number of absolute differences between test and reference device never <35 within 5 mmHg and never <40 within 10 and 15 mmHg. The test device also passed phase 2 of the validation study with a mean (+/-SD) device-observer difference of -1.4+/-4.8 mmHg for systolic and -0.9+/-3.5 mmHg for diastolic blood pressure.
CONCLUSIONS: According to the results of the validation study on the basis of the ESH International Protocol, the Artsana CSI 610 can be recommended for clinical use in adults.

Entities:  

Mesh:

Year:  2007        PMID: 17496468     DOI: 10.1097/MBP.0b013e3280b08394

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  3 in total

1.  Agreement of blood pressure measurements between random-zero and standard mercury sphygmomanometers.

Authors:  Wenjie Yang; Dongfeng Gu; Jing Chen; Cashell E Jaquish; D C Rao; Xigui Wu; James E Hixson; Xiufang Duan; Tanika N Kelly; L Lee Hamm; Paul K Whelton; Jiang He
Journal:  Am J Med Sci       Date:  2008-11       Impact factor: 2.378

2.  A systematic review of statistical methods used to test for reliability of medical instruments measuring continuous variables.

Authors:  Rafdzah Zaki; Awang Bulgiba; Noorhaire Nordin; Noor Azina Ismail
Journal:  Iran J Basic Med Sci       Date:  2013-06       Impact factor: 2.699

3.  Validation of the Oregon Scientific BPU 330 for self-monitoring of blood pressure according to the International Protocol.

Authors:  Li Li; XinYu Zhang; ChunHong Yan; QingXiang Liang
Journal:  Vasc Health Risk Manag       Date:  2008
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.